Visian Toric ICL (Implantable Collamer Lens for Myopia) TMICL 12.6 (-13.5/4.0)Recalls & reported problems
Staar Surgical Company · Diopter: -13.5; Cylinder 4.0; Overall Diameter (Length): 12.6 · Class 3 · Implant
All models of Visian Toric ICL (Implantable Collamer Lens for Myopia) (293)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841542107396), received between Aug 13, 2019 and Apr 20, 2022 — classified as 3 injuries.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size2
- Device Dislodged or Dislocated1
- Misfocusing1
- No Apparent Adverse Event1
- Off-Label Use1
Patient problems
- No Code Available2
- Blurred Vision1
- Intraocular Pressure Increased1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 20, 2022Misfocusing; Device Dislodged or Dislocated; Misfocusing; Device Dislodged or Dislocated
THE REPORTER INDICATED THAT A 12.6MM, TMICL12.6, -13.50/4.0/090 (SPHERE/CYLINDER/AXIS) IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2021. LENS ROTATION NOT ASSOCIATED TO A LOW VAULT AND REFRACTIVE CHANGE OVERTIME WAS OBSERVED. ON (B)(6) 2022 LASIK ENHANCEMENT WAS PERFORMED. ON (B)(6) 2022 THE LENS ROTATED AGAIN, IT WAS REPORTED THAT LENS REPOSITIONING WAS PERFORMED. ON (B)(6) 2022 THE LENS WAS EXCHANGED FOR A SAME LENGTH DIFFERENT MODEL AND POWER LENS. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Blurred Vision
MAUDE report 14164530 ↗InjuryDec 27, 2019Off-Label Use; Inadequacy of Device Shape and/or Size; No Apparent Adverse Event; Off-Label Use; Inadequacy of Device Shape and/or Size; No Apparent Adverse Event
THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.6MM TMICL12.6 IMPLANTABLE COLLAMER LENS, - 13.50/+4.0/092 (SPHERE/CYLINDER/AXIS), IN THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2019. THE LENS WAS REPORTED AS HAVING HIGH VAULTING AND WILL BE EXPLANTED/EXCHANGED FOR A SHORTER LENS. THE LENS REMAINS IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.
Patient problems: No Code Available; No Known Impact Or Consequence To Patient
MAUDE report 9528016 ↗InjuryAug 13, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT A 12.6MM, TMICL12.6, -13.50/4.0/090 (SPHERE/CYLINDER/AXIS), IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2019. PUPIL BLOCK, ANGLE CLOSURE WITH ELEVATED IOP (50 MMHG), AND EXCESSIVE VAULT WERE OBSERVED. ON (B)(6) 2019 THE LENS WAS REMOVED AND EXCHANGED WITH A SMALLER LENGTH LENS. REPORTER NOTED "VAULT IS STILL GENEROUS BUT SAFE." THIS RESOLVED THE PROBLEM. CAUSE OF THE EVENT IS REPORTED AS "RECOMMENDED LENS TOO LARGE." PER REPORTER, PATIENTS CONDITION ON LAST VISIT, (B)(6) 2019, BCVA 20/50, "EYE IS SAFE, UNINFLAMED, GOOD IOP, STILL GENEROUS VAULT BUT SAFE EXAM & IOP, ANGLES OPEN."
Patient problems: Intraocular Pressure Increased; No Code Available
MAUDE report 8889704 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
Other Phakic Toric Intraocular Lens devices
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 13.2 (-13.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-13.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-8.5/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.1 (-5.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-11.5/2.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-14.0/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-5.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-7.5/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page