Skip to content

Visian Toric ICL (Implantable Collamer Lens for Myopia) TMICL 13.2 (-9.5/1.0)Recalls & reported problems

Staar Surgical Company · Diopter: -9.5; Cylinder 1.0; Overall Diameter (Length): 13.2 · Class 3 · Implant

All models of Visian Toric ICL (Implantable Collamer Lens for Myopia) (293)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00841542108942), received between May 14, 2019 and May 10, 2022 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (4 reports).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size4
  • Misfocusing1
  • No Apparent Adverse Event1
  • Off-Label Use1
  • Optical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Blurred Vision1
  • Halo1
  • Intraocular Pressure Increased1
  • No Known Impact Or Consequence To Patient1
  • Visual Disturbances1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 10, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT A 13.2 TMICL 13.2 IMPLANTABLE COLLAMER LENS OF -11.50/1.0/075 (SPHERE/CYLINDER/AXIS) LENS WAS IMPLANTED INTO THE PATIENTS EYE ON AN UNKNOWN DATE. THE LENS WAS REPORTEDLY REMOVED AND REPLACED FOR A SHORTER LENGTH DUE TO THE LENS BEING OVERSIZED. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14350128 ↗
  • InjuryFeb 22, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM TMICL13.2; -9.50/1.0/080 (SPHERE/CYLINDER/AXIS) IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2022. IT WAS REPORTED THERE IS LOW VAULT. LENS REMAINS IMPLANTED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE NOT BEEN SUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13580869 ↗
  • MalfunctionSep 7, 2021Off-Label Use; No Apparent Adverse Event; Off-Label Use; No Apparent Adverse Event

    A 13.2MM TMICL13.2 IMPLANTABLE COLLAMER LENS, 09.50/+1.0/066 (SPHERE/CYLINDER/AXIS), FOR THE LEFT EYE (OS), IS BEING RETURNED TO THE MANUFACTURER AND THE REASON IS UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12437071 ↗
  • InjuryJul 8, 2021Misfocusing; Inadequacy of Device Shape and/or Size; Optical Problem; Misfocusing; Inadequacy of Device Shape and/or Size; Optical Problem

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.2MM TMICL13.2 IMPLANTABLE COLLAMER LENS, -09.50/+1.0/086 (SPHERE/CYLINDER/AXIS), IN THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2021. THE LENS WAS EXPLANTED ON (B)(6) 2021 DUE TO EXCESSIVE VAULTING, ELEVATED INTRAOCULAR PRESSURE (IOP), GLARE/HALOS AND BLURRED VISION. THE LENS WAS EXCHANGED FOR A SHORTER LENS AND THE PROBLEM WAS RESOLVED. GOOD POSITION AND VAULT, SYMPTOMS (GLARE) RESOLVED AND IOP NORMAL. THE CAUSE OF THE EVENT WAS THE DEVICE.

    Patient problems: Blurred Vision; Halo; Intraocular Pressure Increased; Visual Disturbances

    MAUDE report 12140318 ↗
  • MalfunctionMay 14, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM TMICL13.2 LENS, -9.5/+1.0/095 (SPHERE/CYLINDER/AXIS) INTO THE PATIENT'S LEFT (OS) EYE ON (B)(6) 2019. THE SURGEON STATES THIS IS AN UNEXPECTED CASE OF HIGH VAULT. REPORTEDLY, THE LENS REMAINS IMPLANTED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION OR UPDATES HAVE NOT BEEN SUCCESSFUL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8609189 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.

Other Phakic Toric Intraocular Lens devices

All Phakic Toric Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page