Visian Toric ICL (Implantable Collamer Lens for Myopia) TMICL 13.2 (-13.0/2.0)Recalls & reported problems
Staar Surgical Company · Diopter: -13.0; Cylinder 2.0; Overall Diameter (Length): 13.2 · Class 3 · Implant
All models of Visian Toric ICL (Implantable Collamer Lens for Myopia) (293)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841542109451), received between Nov 22, 2019 and May 26, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
- Misfocusing1
Patient problems
- Blurred Vision1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 26, 2021Misfocusing; Device Dislodged or Dislocated; Misfocusing; Device Dislodged or Dislocated
IT WAS REPORTED THAT A 13.2MM, TMICL13.2, -13/2/90 (SPHERE/CYLINDER/AXIS), IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENTS LEFT EYE (0S) ON (B)(6) 2021. ON (B)(6) 2021 LENS REPOSITIONING WAS PERFORMED DUE TO LENS ROTATION (UNKNOWN WHETHER VAULT WAS LOW), IT WAS REPORTED THAT REPOSITIONING RESOLVED THE PROBLEM. FOR STATUS OF THE EYE (SIGN/SYMPTOMS): "RESOLVED" WAS REPORTED. PER STUDY INVESTIGATOR, THE CAUSE OF THIS EVENT WAS RELATED TO THE DEVICE. FOR ADDITIONAL INFORMATION THE REPORTER STATED "PER MEDICAL MONITOR OF NOTE, LENS POWER CALCULATION RECOMMENDED BY STAAR WAS TMICL 13.7, -13.00/2.0/080. HOWEVER, THE DOCTOR CHOSE TO GO WITH THE TMICL 13.2."
Patient problems: Blurred Vision; No Clinical Signs, Symptoms or Conditions
MAUDE report 11891166 ↗InjuryNov 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT HE ATTEMPTED TO INSERT A13.2MM, TMICL13.2, -13.00/2.00/092 (SPHERE/CYLINDER/AXIS) INTO THE PATIENTS RIGHT EYE (OD) TWO TIMES AND THEN SWITCHED TO THE BACKUP LENS. THE REPORTER STATED " WHEN I INITIALLY TRIED TO INSERT THE ICL, IT WAS NOT GLIDING THROUGH THE INCISION AND UNFOLDING PROPERLY. I THOUGHT THAT MAY BE DUE TO MY INCISION BEING TOO SMALL SO I THEN I ENLARGED MY INCISION SLIGHTLY. WHEN I TRIED TO INSERT THE ICL AGAIN THE PLUNGER STUCK TO THE ICL, I DIDN'T KNOW IF IT WAS AN ISSUE WITH THE ICL, PLUNGER, CARTRIDGE, OR INJECTOR SO I SWITCHED EVERYTHING OUT IN ORDER TO ENSURE I HAD AN ICL WITH PROPER INTEGRITY FOR THE PATIENT." THE ICL TOUCHED THE PATIENTS EYE BUT WAS NEVER FULLY INSERTED INTO THE PATIENTS EYE. IT WAS ONLY PARTIALLY INSERTED. THE BACKUP LENS WAS USED WITH NO COMPLICATIONS.
Patient problems: No Code Available
MAUDE report 9362222 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
Other Phakic Toric Intraocular Lens devices
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 13.2 (-13.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-13.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-8.5/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.1 (-5.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-11.5/2.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-14.0/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-5.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-7.5/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.