Skip to content

Visian Toric ICL (Implantable Collamer Lens for Myopia) TMICL 13.7 (-16.0/4.0)Recalls & reported problems

Staar Surgical Company · Diopter: -16.0; Cylinder 4.0; Overall Diameter (Length): 13.7 · Class 3 · Implant

All models of Visian Toric ICL (Implantable Collamer Lens for Myopia) (293)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841542112086), received between Mar 18, 2021 and Mar 18, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method2
  • Inadequacy of Device Shape and/or Size2

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Intraocular Pressure Increased2
  • Pupillary Block2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2021Inadequacy of Device Shape and/or Size; Improper or Incorrect Procedure or Method; Inadequacy of Device Shape and/or Size; Improper or Incorrect Procedure or Method

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.7MM TMICL13.7, -16.00/4.0/065 (SPHERE/CYLINDER/AXIS), IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2021. THE LENS WAS EXCHANGED WITH A FOR A SHORTER LENGTH LENS DUE TO PUBIL BLOCK WITH ELEVATED IOP(34MMHG), ANGLE CLOSURE WITH ELEVATED IOP((34MMHG) ASSESSED ON (B)(6) 2021), AND EXCESSIVE VAULT, ON (B)(6) 2021. THIS RESOLVED THE PROBLEM. CAUSE OF THE EVENT IS REPORTED AS USER ERROR.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Intraocular Pressure Increased; Pupillary Block

    MAUDE report 11515299 ↗
  • InjuryMar 18, 2021Inadequacy of Device Shape and/or Size; Improper or Incorrect Procedure or Method; Inadequacy of Device Shape and/or Size; Improper or Incorrect Procedure or Method

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.7MM TMICL13.7, -16.00/4.0/125 (SPHERE/CYLINDER/AXIS), IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2021. THE LENS WAS EXCHANGED WITH A FOR A SHORTER LENGTH LENS DUE TO PUBIL BLOCK WITH ELEVATED IOP(39MMHG), ANGLE CLOSURE WITH ELEVATED IOP(39MMHG) ASSESSED ON (B)(6) 2021, AND EXCESSIVE VAULT, ON (B)(6) 2020. THIS RESOLVED THE PROBLEM. CAUSE OF THE EVENT IS REPORTED AS USER ERROR.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Intraocular Pressure Increased; Pupillary Block

    MAUDE report 11515298 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.

Other Phakic Toric Intraocular Lens devices

All Phakic Toric Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page