Skip to content

EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-11.5/2.0) (REF VTICM5 12.6)Recalls & reported problems

Staar Surgical Company · VTICM5 - Diopter: -11.5; Cylinder 2.0; Overall Diameter (Length): 12.6 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841542115193), received between Nov 15, 2022 and Nov 14, 2023 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Unintended Movement1

Patient problems

  • Intraocular Pressure Increased2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Endophthalmitis1
  • Hypopyon1
  • Inflammation1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.6MM VTICM5_12.6 IMPLANTABLE COLLAMER LENS, -11.50/+2.0/091 (SPHERE/CYLINDER/AXIS), INTO THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2023. THE PATIENT WAS FOUND TO HAVE ENDOPHTHALMITIS AND PERMANENT LOSS OF BCVA. INTRAOCULAR PRESSURE (IOP) CHECK AND ANTERIOR SEGMENT IMAGING WERE PERFORMED ON (B)(6) 2023. PATIENT WAS PRESCRIBED MEDICATIONS FOR INCREASED IOP. PATIENT HAD HYPOPYON, FIBRIM NET, CELLS IN AC, WITH REDNESS, SEVERE LIGHT SENSITIVITY AND PAIN, INITIALLY DIAGNOSED AS TASS DUE TO FAST COURSE OF SIGNS AND CLEAR B-SCAN BY DOCTOR AND RETINAL SPECIALIST. PATIENT HAD VITREOUS BIOPSY WITH INTRAVITREAL INJECTION OF VANCOMYCIN AND CEFTAZIDIME TO RULE OUT AND TREAT ENDOPHTHALMITIS AS WELL AS SUBTENON KENALOG DUE TO INITIAL FAST COURSE OF SIGNS OF TASS. THE NEXT DAY, DUE TO CELLS ON VITREOUS B-SCAN, THE PATIENT HAD BILATERAL PPV (PARS PLANA VITRECTOMY) WITH INTRAVITREAL INJECTION OF CEFTAZIDIME, VANCOMYCIN AND VITREOUS CULTURE. VITREOUS CULTURE GREW PSEUDOMONA AERUGINOSA AND CURRENTLY THE PATIENT IS HOSPITALIZED. THE LENS REMAINED IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Endophthalmitis; Hypopyon; Inflammation; Intraocular Pressure Increased; Pain

    MAUDE report 18136578 ↗
  • MalfunctionJun 5, 2023Unintended Movement; Unintended Movement

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED UPSIDE DOWN A 12.6MM VTICM5_12.6, -11.5/2.0/110 (SPHERE/CYLINDER/AXIS), IMPLANTABLE COLLAMER LENS, INTO THE PATIENT'S RIGHT EYE(OD) ON (B)(6) 2023. THERE WAS PATIENT CONTACT (LENS TOUCHED THE EYE) WITH NO PATIENT INJURY. THE LENS WAS REMOVED AND TORE OUTSIDE THE PATIENTS EYE AND BACK UP LENS WAS IMPLANTED INTRAOPERATIVLY ON (B)(6) 2023. THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS REPORTED AS DEVICE, UNFOLDED UPSIDE DOWN AND TORE OUTSIDE EYE. REPORTEDLY PATIENT REMAINS IN STUDY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17066560 ↗
  • InjuryNov 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.6MM VTICM5_12.6 IMPLANTABLE COLLAMER LENS, -11.50/+2.0/093 (SPHERE/CYLINDER/AXIS), IN THE PATIENTS LEFT EYE (OS) ON (B)(6) 2022. THE INTRAOCULAR PRESSURE (IOP) WAS REPORTED AS BEING ELEVATED AND COMBIGAN BID AND PRED FORTE BID WERE INITIATED. THE ELEVATED IOP WAS RESOLVED AND WAS MEASURED WITHIN NORMAL RANGE ON (B)(6) 2022. THE LENS REMAINED IMPLANTED.

    Patient problems: Intraocular Pressure Increased

    MAUDE report 15801805 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.

Other Phakic Toric Intraocular Lens devices

All Phakic Toric Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page