Skip to content

EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 13.2 (-8.5/2.5) (REF VTICM5 13.2)Recalls & reported problems

Staar Surgical Company · VTICM5 - Diopter: -8.5; Cylinder 2.5; Overall Diameter (Length): 13.2 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841542116398), received between Mar 14, 2024 and Mar 19, 2026 — classified as 1 injury, 1 malfunction.

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • No Apparent Adverse Event1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2026Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED LENS ROTATION. THE LENS WAS REPOSITIONED. CAUSE OF THE EVENT IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24641563 ↗
  • MalfunctionMar 14, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM VTICM513.2 IMPLANTABLE COLLAMER LENS OF A -8.5/2.5/051 (SPHERE/CYLINDER/AXIS) INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2024. IT IS REPORTED THAT THE LENS MAY HAVE A HIGH VAULT AND POSSIBILY ROTATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18909479 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.

Other Phakic Toric Intraocular Lens devices

All Phakic Toric Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page