EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5) (REF VICM5 12.6)Recalls & reported problems

Staar Surgical Company · VICM5 - Diopter: -9.5; Overall Diameter (Length): 12.6 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00841542119177), received between Dec 7, 2022 and Mar 26, 2024 — classified as 5 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

7 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 29, 2023Class IIOpen, Classified

    Their is a potential that intraocular implant devices may not meet specifications.

    Z-0465-2024 · Root cause: Software design (manufacturing process) · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size4
  • No Apparent Adverse Event3
  • Misfocusing2
  • Off-Label Use1
  • Optical Problem1
  • Patient-Device Incompatibility1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Halo2
  • Intraocular Pressure Increased2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Blurred Vision1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT A 12.6MM VICM5 12.6 IMPLANTABLE COLLAMER LENS OF -9.50 DIOPTER LENS TORE DURING LOADING. THE LENS HAD NO PATIENT CONTACT. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18981679 ↗
  • MalfunctionDec 27, 2023Unintended Movement; No Apparent Adverse Event; Unintended Movement; No Apparent Adverse Event

    THE REPORTER INDICATED THAT 12.6MM VICM512.6 IMPLANTABLE COLLAMER LENS OF A -9.5 DIOPTER WAS IMPLANTED UPSIDED DOWN. THE LENS WAS REMOVED AND REPLACEMENT LENS WAS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18406812 ↗
  • InjuryOct 23, 2023Inadequacy of Device Shape and/or Size; Patient-Device Incompatibility; Inadequacy of Device Shape and/or Size; Patient-Device Incompatibility

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5 12.6, -9.50 DIOPTER, IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON 31/AUG/2023. LOW VAULT WAS OBSERVED. THE LENS REMAINS IMPLANTED. CAUSE OF THE EVENT IS REPORTED AS PATIENT RELATED FACTOR. DEVICE DID NOT FAIL TO PERFORM AS INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17986862 ↗
  • InjurySep 28, 2023Misfocusing; Optical Problem; Misfocusing; Optical Problem

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.6MM VICM5_12.6 IMPLANTABLE COLLAMER LENS, -09.50 DIOPTER, INTO THE PATIENTS EYE ON (B)(6) 2023. THE LENS WAS REMOVED ON (B)(6) 2023 DUE TO OVER CORRECTION. THE LENS WAS REPLACED WITH A SHORTER LENS. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.

    Patient problems: Halo; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17831218 ↗
  • InjuryJul 7, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM512.6, -09.50 DIOPTER, IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2023. THE LENS WAS REPLACED WITH A SHORTER LENGTH LENS DUE TO SIGNIFICANT REDUCTION OF IRIDOCORNEAL ANGLES, ANGLE CLOSURE WITH ELEVATED IOP(41MMHG, ASSESSED ON (B)(6) 2023), AND EXCESSIVE VAULT ON (B)(6) 2023. THIS RESOLVED THE PROBLEM. CAUSE OF THE EVENT IS REPORTED AS DEVICE, DIAMETER WAS TOO LARGE, REPLACED SMALLER LENS FIXING PROBLEM.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Intraocular Pressure Increased

    MAUDE report 17279000 ↗
  • MalfunctionApr 25, 2023No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT THERE IS A POSSIBLE ISSUE FOR A 12.6MM VICM512.6 IMPLANTABLE COLLAMER LENS. NO FURTHER INFORMATION IS ABLE TO BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16809010 ↗
  • InjuryFeb 6, 2023Misfocusing; Off-Label Use; Inadequacy of Device Shape and/or Size; Misfocusing; Off-Label Use; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5_12.6; -9.5 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2022. THE PATIENT EXPERIENCED LOW VAULTING WITH ELEVATED INTRAOCULAR PRESSURE; GLARE/HALOES; BLURRED VISION. ON (B)(6) 2023 THE LENS WAS EXPLANTED. ON THIS SAME DATE IN A SEPARATE SURGERY A REPLACEMENT LENS OF LONGER LENGTH WAS IMPLANTED; IT WAS ADDITIONALLY INDICATED AN INCISION ENLARGEMENT AND ADDITION SUTURE WAS PERFORMED. THIS RESOLVED THE PROBLEM. THE CAUSE OF THE EVENT WAS REPORTED AS THE DEVICE AND NOTED "SHALLOW VAULT DUE TO LENGTH".

    Patient problems: Blurred Vision; Halo; Intraocular Pressure Increased

    MAUDE report 16309075 ↗
  • InjuryDec 7, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5_12.6; -9.50 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS). IT WAS REPORTED THAT THE LENS WAS EXPLANTED DUE TO OVER SIZED VAULTING. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE NOT BEEN SUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15936590 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.

Other Lens, Intraocular, Phakic devices

All Lens, Intraocular, Phakic devices →