EVO/EVO+VISIAN Implantable Collamer Lens VTICMO 13.2 (-15.0/1.5) (REF VTICMO 13.2)Recalls & reported problems
Staar Surgical Company · VTICMO - Diopter: -15.0; Cylinder 1.5; Overall Diameter (Length): 13.2 · Class 3 · Implant
All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841542120937), received between Nov 23, 2022 and Nov 23, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size2
- Misfocusing2
- Off-Label Use2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Blurred Vision2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 23, 2022Misfocusing; Off-Label Use; Inadequacy of Device Shape and/or Size; Misfocusing; Off-Label Use; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.2MM VTICMO13.2 IMPLANTABLE COLLAMER LENS, -15.00/+1.5/072 (SPHERE/CYLINDER/AXIS), IN THE PATIENTS LEFT EYE (OS) ON (B)(6) 2022. THE LENS WAS EXPLANTED ON (B)(6) 2022 DUE TO EXCESSIVE VAULTING, SIGNIFICANT REDUCTION OF IRIDO-CORNEAL ANGLES AND BLURRED VISION. THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT WAS DUE TO THE DEVICE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision
MAUDE report 15856830 ↗InjuryNov 23, 2022Misfocusing; Off-Label Use; Inadequacy of Device Shape and/or Size; Misfocusing; Off-Label Use; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.2MM VTICMO13.2 IMPLANTABLE COLLAMER LENS, -15.00/+1.5/106 (SPHERE/CYLINDER/AXIS), IN THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2022. THE LENS WAS EXPLANTED ON (B)(6) 2022 DUE TO EXCESSIVE VAULTING, SIGNIFICANT REDUCTION OF IRIDO-CORNEAL ANGLES AND BLURRED VISION. THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT WAS DUE TO THE DEVICE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision
MAUDE report 15856726 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
Other Phakic Toric Intraocular Lens devices
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 13.2 (-13.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-13.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-8.5/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.1 (-5.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-11.5/2.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-14.0/2.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-5.5/1.5)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.6 (-7.5/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page