Skip to content

Ophthalmic Spatula K3-2538Recalls & reported problems

Katena Products, Inc. · INTRALASE FLAP LIFTER DOUBLE ENDED · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841668103210), received between Jan 24, 2024 and Jan 24, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • Discomfort1
  • Visual Disturbances1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 24, 2024Material Fragmentation; Material Fragmentation

    ON 26-DEC-2023, A DISTRIBUTOR FROM FRANCE INFORMED CORZA MEDICAL CUSTOMER SERVICE REPRESENTATIVE THAT THEY RECEIVED A COMPLAINT FROM THEIR CUSTOMER AGAINST K3-2538 INTRALASE FLAP LIFTER DOUBLE. THE DISTRIBUTOR STATED THAT ON (B)(6) 2022, THE CUSTOMER PURCHASED 6 MANIPULATORS. THESE MANIPULATORS ARE USED BY A SINGLE DOCTOR WHO PERFORMS MULTIPLE PROCEDURES ONCE A MONTH. OVER SEVERAL MONTHS, THE SURGEON NOTICED METAL DEPOSITS IN THE POSTOPERATIVE INTERFACE. THE DATES OF THESE PROCEDURES WERE ON (B)(6) 2023. THERE ARE MULTIPLE INSTRUMENTS INVOLVED DURING THE PROCEDURES INCLUDING A NON-KATENA MICROMANIPULATOR AND A BLEPHAROSTAT. THE DOCTOR STATED THAT ONE PATIENT EXPERIENCED VISUAL DISCOMFORT AND PHOTIC EFFECTS. THE 6 KATENA K3-2538 INSTRUMENTS WERE REMOVED FROM FUTURE PROCEDURES AND SENT TO KATENA FOR EVALUATION. ON 22-JAN-2023 KATENA (CORZA OPTHALMOLOGY) RECEIVED THE INSTRUMENTS.

    Patient problems: Discomfort; Visual Disturbances

    MAUDE report 18576893 ↗

Other Spatula, ophthalmic devices

All Spatula, ophthalmic devices →

Sources and updates