Ophthalmic Spatula K3-2538Recalls & reported problems
Katena Products, Inc. · INTRALASE FLAP LIFTER DOUBLE ENDED · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841668103210), received between Jan 24, 2024 and Jan 24, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation1
Patient problems
- Discomfort1
- Visual Disturbances1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 24, 2024Material Fragmentation; Material Fragmentation
ON 26-DEC-2023, A DISTRIBUTOR FROM FRANCE INFORMED CORZA MEDICAL CUSTOMER SERVICE REPRESENTATIVE THAT THEY RECEIVED A COMPLAINT FROM THEIR CUSTOMER AGAINST K3-2538 INTRALASE FLAP LIFTER DOUBLE. THE DISTRIBUTOR STATED THAT ON (B)(6) 2022, THE CUSTOMER PURCHASED 6 MANIPULATORS. THESE MANIPULATORS ARE USED BY A SINGLE DOCTOR WHO PERFORMS MULTIPLE PROCEDURES ONCE A MONTH. OVER SEVERAL MONTHS, THE SURGEON NOTICED METAL DEPOSITS IN THE POSTOPERATIVE INTERFACE. THE DATES OF THESE PROCEDURES WERE ON (B)(6) 2023. THERE ARE MULTIPLE INSTRUMENTS INVOLVED DURING THE PROCEDURES INCLUDING A NON-KATENA MICROMANIPULATOR AND A BLEPHAROSTAT. THE DOCTOR STATED THAT ONE PATIENT EXPERIENCED VISUAL DISCOMFORT AND PHOTIC EFFECTS. THE 6 KATENA K3-2538 INSTRUMENTS WERE REMOVED FROM FUTURE PROCEDURES AND SENT TO KATENA FOR EVALUATION. ON 22-JAN-2023 KATENA (CORZA OPTHALMOLOGY) RECEIVED THE INSTRUMENTS.
Patient problems: Discomfort; Visual Disturbances
MAUDE report 18576893 ↗
Other Spatula, ophthalmic devices
- Oph.Spatula 05-4036Katena Products, Inc. · Spatula, ophthalmic1 report
- Oph.Spatula 08-14527Katena Products, Inc. · Spatula, ophthalmic1 report
- Oph.Spatula 08-14502Katena Products, Inc. · Spatula, ophthalmic1 report
- KOCH STOP CHOP MANIPUL 12B SUF01Bausch & Lomb GmbH · Spatula, Ophthalmic1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page