OMNI Knee KC-35041Recalls & reported problems
Omni Life Science, Inc. · Congruent Tibial Insert, Size 4 x 11mm · Class 2 · Implant
All models of OMNI Knee (78)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00841690101314), received between Nov 29, 2017 and Jan 23, 2025 — classified as 10 injuries.
The most frequently reported specific device problem is “Unstable” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable3
- Adverse Event Without Identified Device or Use Problem2
- Appropriate Device Problem Term/Code Not Available2
- Patient Device Interaction Problem2
- Insufficient Device Problem Information1
Patient problems
- Unspecified Infection5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Fall1
- Joint Laxity1
- No Code Available1
- Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJan 23, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2025. THE ORIGINAL SURGERY DATE IS (B)(6) 2020. THE REASON FOR REVISION IS REPORTED PATIENT PAIN. DURING THE REVISION, DURING THE REVISION, THE ORIGINAL TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW IMPLANTS.
Patient problems: Pain
MAUDE report 21218804 ↗InjurySep 19, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2024. THE ORIGINAL SURGERY DATE IS (B)(6) 2016. THE REASON FOR REVISION IS REPORTED INSTABILITY. DURING THE REVISION THE INSERT AND BOLT WERE REMOVED AND REPLACED WITH NEW IMPLANTS.
Patient problems: Joint Laxity
MAUDE report 20255885 ↗InjuryNov 4, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2022. THE ORIGINAL SURGERY DATE IS (B)(6) 2022. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION, INSERT AND BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.
Patient problems: Unspecified Infection
MAUDE report 15736515 ↗InjuryJul 11, 2022Appropriate Device Problem Term/Code Not Available
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6)2022. THE ORIGINAL SURGERY DATE IS (B)(6) 2021. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION, THE FEMUR, INSERT AND BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.
Patient problems: Unspecified Infection
MAUDE report 14980998 ↗InjuryAug 20, 2021Adverse Event Without Identified Device or Use Problem
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2021. THE ORIGINAL SURGERY DATE IS (B)(6) 2020. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION, TIBIAL INSERT AND RETAINING BOLT, WERE REMOVED AND REPLACED WITH NEW COMPONENTS.
Patient problems: Unspecified Infection
MAUDE report 12348529 ↗InjuryAug 18, 2021Appropriate Device Problem Term/Code Not Available
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2021. THE ORIGINAL SURGERY DATE IS (B)(6) 2020. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION, TIBIAL INSERT AND RETAINING BOLT, BASEPLATE, PATELLA, FEMUR WERE REMOVED AND REPLACED WITH NEW NON OMNI COMPONENTS.
Patient problems: Unspecified Infection
MAUDE report 12339325 ↗InjuryNov 19, 2020Unstable
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2020. THE ORIGINAL SURGERY DATE IS (B)(6) 2016. THE REASON FOR REVISION IS REPORTED INSTABILITY. DURING THE REVISION, THE ORIGINAL TIBIAL INSERT, PATELLA AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 10872830 ↗InjuryAug 27, 2018Unstable
THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2018. THE ORIGINAL SURGERY IS DATED (B)(6) 2017. THE REVISION SURGERY OCCURRED BECAUSE OF INSTABILITY DUE TO A REPORTED PATIENT FALL. DURING THE REVISION, THE ORIGINAL OMNI SIZE 4 X 11MM CONGRUENT INSERT AND TIBIAL BASEPLATE RETAINING BOLT WERE REMOVED AND REPLACED WITH A SIZE 4 X 12MM CONGRUENT INSERT AND ANOTHER TIBIAL BASEPLATE RETAINING BOLT.
Patient problems: Fall
MAUDE report 7817827 ↗InjuryJun 6, 2018Adverse Event Without Identified Device or Use Problem
THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2018. THE ORIGINAL SURGERY IS DATED (B)(6) 2018. THE REVISION SURGERY OCCURRED BECAUSE OF AN INFECTION, AS REPORTED. DURING THE REVISION, THE ORIGINAL OMNI DOVETAIL INSERT LOCKING CONNECTOR AND DOVETAIL CONGRUENT INSERT, SZ 4, 11MM WAS REMOVED AND REPLACED WITH NEW COMPONENTS OF THE SAME SIZE.
Patient problems: Unspecified Infection
MAUDE report 7575331 ↗InjuryNov 29, 2017Unstable
THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2017. THE ORIGINAL SURGERY IS DATED (B)(6) 2015. THE REVISION SURGERY OCCURRED BECAUSE OF INSTABILITY. DURING THIS REVISION, THE ORIGINAL OMNI TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND EXCHANGED FOR THICKER COMPONENTS.
Patient problems: No Code Available
MAUDE report 7069158 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page