OMNI Knee KC-36050Recalls & reported problems

Omni Life Science, Inc. · Ultra Tibial Insert CS, Size 5 x 10mm · Class 2 · Implant

All models of OMNI Knee (78)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00841690101734), received between Nov 10, 2017 and May 5, 2025 — classified as 20 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information9
  • Adverse Event Without Identified Device or Use Problem7
  • Appropriate Device Problem Term/Code Not Available4

Patient problems

  • Unspecified Infection14
  • Pain3
  • Fall1
  • Joint Laxity1
  • Loss of Range of Motion1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2025. THE ORIGINAL SURGERY DATE (B)(6) 2023. THE REASON FOR REVISION IS REPORTED INFECTION. DURING THE REVISION THE TIBIAL INSERT AND BOLT WERE REMOVED AND REPLACED ANTIBIOTIC SPACERS.

    Patient problems: Unspecified Infection

    MAUDE report 21959746 ↗
  • InjuryJan 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2023. THE ORIGINAL SURGERY DATE IS (B)(6) 2023. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION THE INSERT AND BOLT WERE REMOVED AND REPLACED WITH NEW IMPLANTS.

    Patient problems: Unspecified Infection

    MAUDE report 18489914 ↗
  • InjuryOct 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2023. THE ORIGINAL SURGERY DATE IS (B)(6) 2023. THE REASON FOR REVISION IS REPORTED INSTABILITY. DURING THE REVISION, INSERT AND BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Joint Laxity

    MAUDE report 18020014 ↗
  • InjuryOct 14, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6)2022. THE ORIGINAL SURGERY DATE 10/08/2019. THE REASON FOR REVISION IS REPORTED PATIENT PAIN. DURING THE REVISION SURGEON REMOVED THE TIBIAL INSERT AND BOLT REPLACED WITH NEW IMPLANT.

    Patient problems: Pain

    MAUDE report 15610370 ↗
  • InjurySep 13, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2022. THE ORIGINAL SURGERY DATE IS (B)(6) 2022. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION, INSERT AND BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Fall

    MAUDE report 15414572 ↗
  • InjuryJul 6, 2022Appropriate Device Problem Term/Code Not Available

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2022. THE ORIGINAL SURGERY DATE IS (B)(6) 2022. THE REASON FOR REVISION IS INFECTION. DURING THE REVISION, TIBIAL INSERT AND RETAINING BOLT, WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Unspecified Infection

    MAUDE report 14939900 ↗
  • InjuryJun 17, 2021Adverse Event Without Identified Device or Use Problem

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2021. THE ORIGINAL SURGERY DATE IS (B)(6) 2019. THE REASON FOR REVISION IS SUSPECTED INFECTION. DURING THE REVISION, THE FEMORAL COMPONENT AND TIBIAL INSERT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Unspecified Infection

    MAUDE report 12019890 ↗
  • InjuryMar 18, 2021Appropriate Device Problem Term/Code Not Available

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2021. THE ORIGINAL SURGERY DATE IS (B)(6) 2020. THE REASON FOR REVISION IS SUSPECTED INFECTION. DURING THE REVISION, THE TIBIAL INSERT AND RETAINING BOLT WAS REMOVED AND REPLACED WITH NEW REVISION COMPONENTS.

    Patient problems: Unspecified Infection

    MAUDE report 11521456 ↗
  • InjuryJul 17, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2020. THE ORIGINAL SURGERY IS (B)(6) 2017. THE REASON FOR REVISION IS REPORTED INFECTION. DURING THE REVISION, THEY REMOVED THE ORIGINAL TIBIAL INSERT AND LOCKING BOLT AND REPLACED THEM WITH NEW COMPONENTS.

    Patient problems: Unspecified Infection

    MAUDE report 10290733 ↗
  • InjuryJun 24, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2020. THE ORIGINAL SURGERY DATE IS (B)(6) 2020. THE REASON FOR REVISION IS REPORTED PATIENT PAIN. DURING THE REVISION, SURGEON REPLACE TIBIA INSERT AND RETAINING BOLT.

    Patient problems: Pain

    MAUDE report 10191693 ↗
  • InjuryMar 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2020. THE ORIGINAL SURGERY IS DATED (B)(6) 2019. THE REASON FOR REVISION IS SUSPECTED INFECTION. DURING THE REVISION, ALL COMPONENTS WERE REMOVED AND AN ANTIBIOTIC SPACER WAS PUT IN WITH PLANS TO COMPLETE REVISION AT A LATER DATE.

    Patient problems: Unspecified Infection

    MAUDE report 9825903 ↗
  • InjuryJan 13, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2019. THE ORIGINAL SURGERY IS DATED (B)(6) 2018. THE SURGERY OCCURED BECAUSE OF REPORTED INFECTION. DURING THE REVISION, THE ORIGINAL TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Unspecified Infection

    MAUDE report 9582517 ↗
  • InjuryDec 13, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2019. THE ORIGINAL SURGERY IS DATED (B)(6) 2018. THE REASON FOR REVISION IS REPORTED INFECTION. DURING THE REVISION, THE ORIGINAL TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED.

    Patient problems: Unspecified Infection

    MAUDE report 9469610 ↗
  • InjuryOct 28, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2019. THE ORIGINAL SURGERY IS DATED (B)(6) 2018. THE REASON FOR REVISION IS REPORTED POOR RANGE OF MOTION. DURING THE REVISION, THE TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Loss of Range of Motion; Patient Problem/Medical Problem

    MAUDE report 9247750 ↗
  • InjurySep 18, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2019. THE ORIGINAL SURGERY IS DATED (B)(6) 2013. THE REASON FOR REVISION IS REPORTED PATIENT PAIN AND INSTABILITY. DURING THE REVISION, THE ORIGINAL TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS OF A LARGER SIZE.

    Patient problems: Pain

    MAUDE report 9085098 ↗
  • InjuryJul 3, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2019. THE ORIGINAL SURGERY IS DATED (B)(6) 2019. THE REVISION SURGERY OCCURED BECAUSE OF REPORTED PATIENT INFECTION. DURING THE REVISION, THE ORIGINAL TIBIAL INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS OF THE SAME SIZE.

    Patient problems: Unspecified Infection

    MAUDE report 8757600 ↗
  • InjuryJul 6, 2018Appropriate Device Problem Term/Code Not Available

    THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2018. THE ORIGINAL SURGERY IS DATED (B)(6) 2011 AND THE FIRST REVISION IS DATED (B)(6) 2018. THIS REVISION SURGERY OCCURRED BECAUSE OF A REPORTED INFECTION. DURING THE REVISION, THE ORIGINAL OMNI SIZE 5 LEFT CR FEMORAL COMPONENT AND SIZE 5X10MM TIBIAL INSERT FROM THE FIRST REVISION WERE REMOVED AND REPLACED WITH COMPONENTS OF THE SAME SIZE.

    Patient problems: Unspecified Infection

    MAUDE report 7667446 ↗
  • InjuryFeb 7, 2018Adverse Event Without Identified Device or Use Problem

    THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2018. THE ORIGINAL SURGERY IS DATED (B)(6) 2018. THE REVISION SURGERY OCCURRED BECAUSE OF INFECTION. DURING THE REVISION, THE OMNI INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS.

    Patient problems: Unspecified Infection

    MAUDE report 7249483 ↗
  • InjuryJan 19, 2018Adverse Event Without Identified Device or Use Problem

    THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2017. THE ORIGINAL SURGERY IS DATED (B)(6) 2017. THE REVISION SURGERY OCCURRED BECAUSE OF INFECTION. DURING THE REVISION, THE ORIGINAL OMNI SIZE 5 X 10MM INSERT AND RETAINING BOLT WERE REMOVED AND REPLACED WITH NEW COMPONENTS OF THE SAME SIZE.

    Patient problems: Unspecified Infection

    MAUDE report 7207832 ↗
  • InjuryNov 10, 2017Appropriate Device Problem Term/Code Not Available

    THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A KNEE REVISION SURGERY ON (B)(6) 2017. THE ORIGINAL SURGERY IS DATED (B)(6) 2015. THE REVISION SURGERY OCCURRED BECAUSE OF INFECTION. DURING THIS REVISION, THE ORIGINAL OMNI SIZE 5 X 10 MM TIBIAL INSERT AND RETAINING BOLT WERE EXCHANGED FOR NEW IMPLANTS OF THE SAME SIZE.

    Patient problems: Unspecified Infection

    MAUDE report 7022591 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →