OMNI ARC Hip HC-22002Recalls & reported problems
Omni Life Science, Inc. · ARC Neck 8° Short · Class 2 · Implant
All models of OMNI ARC Hip (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841690115915), received between Aug 22, 2017 and Sep 18, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Metal Shedding Debris1
- Patient-Device Incompatibility1
Patient problems
- Autoimmune Reaction1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
A COMPLAINT WAS INITIATED FOR A PATIENT WHO UNDERWENT A HIP REVISION SURGERY ON (B)(6) 2014. THE ORIGINAL SURGERY IS DATED (B)(6) 2013. THE REASON FOR REVISION IS REPORTED PATIENT PAIN. DURING THE REVISION, THE ORIGINAL MODULATE NECK AND FEMORAL HEAD WERE REMOVED AND REPLACED WITH NEW COMPONENTS.
Patient problems: Pain
MAUDE report 9085100 ↗InjuryAug 22, 2017Metal Shedding Debris; Patient-Device Incompatibility
THE COMPLAINT INVOLVED A PATIENT WHO UNDERWENT A HIP REVISION SURGERY ON (B)(6)2016. THE ORIGINAL SURGERY IS DATED (B)(6)2012. THE REVISION SURGERY OCCURRED BECAUSE METALLOSIS. DURING THE REVISION, THE OMNI ARC STEM AND ARC NECK WERE REMOVED AND REPLACED WITH NEW COMPONENTS.
Patient problems: Autoimmune Reaction
MAUDE report 6811790 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page