Skip to content

Midmark 153901-003Recalls & reported problems

Midmark Corporation · The Midmark Elevance and Procenter LR Mounted Dental LED Light is intended to be used as a component in a dental environment to provide lighting during dental procedures. · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841709107344), received between Feb 14, 2025 and Feb 14, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device Fell” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Fell1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 14, 2025Device Fell; Device Fell

    IT WAS REPORTED TO MIDMARK CORPORATION THAT ON OR AROUND (B)(6) 2025, A LED DENTAL LIGHT FELL AND HURT A STUDENT AT A DENTAL COLLEGE. THE STUDENT RECEIVED A PINCH FROM THE LIGHT. NO FURTHER DETAILS OF THE SEVERITY OF THE INJURY HAS BEEN MADE AT THE TIME OF THE INITIAL REPORT. DUE TO PREVIOUS REPORTING OF LIKE INSTANCES OF LED DENTAL LIGHTS FALLING, MIDMARK CORPORATION COMPLIED TO REPORT THIS INSTANCE.

    Patient problems: Insufficient Information

    MAUDE report 21389193 ↗

FDA clearance and classification

The FDA regulates product code EAZ (light, operating, dental) under 21 CFR 872.4630 as a Class I device. 21 CFR 872.4630 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Midmark 153901-003

Has the Midmark 153901-003 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00841709107344), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Midmark 153901-003?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 14, 2025. That report lists device problem “Device Fell” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Midmark 153901-003?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Midmark 153901-003 FDA approved?

Devices of this type (product code EAZ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Midmark Corporation.

Who makes the Midmark 153901-003, and is it still sold?

The FDA's device database lists Midmark Corporation as the company responsible for it (brand Midmark, model 153901-003). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other LIGHT, OPERATING, DENTAL devices

All LIGHT, OPERATING, DENTAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page