Skip to content

Vapotherm PF-DPC-LOW-BAG-MXRecalls & reported problems

Vapotherm, Inc. · Precision Flow Disposable Patient Circuit, Low Mexico Made · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841737104971), received between Sep 9, 2024 and Sep 9, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Degraded” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 16, 2022Class IIOpen, Classified

    There have been reports of excessive condensation associated with certain lots of the Low Flow Disposable Patient Circuits (DPCs) used with Precision Flow System which may lead to excessive rainout.

    Z-1272-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Degraded1
  • Fluid/Blood Leak1

Patient problems

  • Apnea1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2024Degraded; Fluid/Blood Leak; Degraded; Fluid/Blood Leak

    THIS INCIDENT HAPPENED IN THE UK AND IS BEING REPORTED PER THE GUIDANCE PROVIDED BY FDA ON 21 CFR 803. THIS INCIDENT WAS ON VAPOTHERM'S PRECISION FLOW SYSTEM WHICH IS AN OPEN SYSTEM WITH A NASAL CANNULA TO TREAT PATIENTS IN RESPIRATORY DISTRESS. CUSTOMER STATED THAT INFANT/NEONATAL PATIENT WAS BEING TREATED WITH VAPOTHERM HIGH VELOCITY THERAPY ON THE PRECISION FLOW DEVICE. THE PATIENT WAS BEING TREATED FOR LONG TERM ON THE DEVICE AND WAS IN THE PROCESS OF BEING WEANED OFF THE THERAPY. EXCESSIVE CONDENSATION WAS OBSERVED TO BE EXITING THE DEVICE AND SPURTING FROM THE PATIENT'S NOSE WHILE BEING MONITORED. PATIENT SUBSEQUENTLY BECAME APNOEIC. PATIENT WAS RESUSCITATED WITH NEOPUFF AND RECOVERED QUICKLY. PATIENT WAS BEING FURTHER MONITORED.

    Patient problems: Apnea

    MAUDE report 20176497 ↗

Other Humidifier, Respiratory Gas, (Direct Patient Interface) devices