Skip to content

Spencer Probe Depth Electrode SD06R-AP58X-000Recalls & reported problems

Ad-Tech Medical Instrument Corporation · Class 2

All models of Spencer Probe Depth Electrode (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 19, 2024Class IICompleted

    Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

    Z-0220-2025 · Root cause: Employee error · Recall details · FDA record ↗

Other ELECTRODE, DEPTH devices

All ELECTRODE, DEPTH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page