Mammotome MAMMOMARK 2 MAM3008Recalls & reported problems
Devicor Medical Products, Inc. · MammoMARK Biopsy Site Identifier 8G U Shape- Non CE · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841911101659), received between Feb 6, 2020 and Nov 2, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Use of Device Problem2
- Mechanical Problem1
Patient problems
- Foreign Body In Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 2, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY THE REPRESENTATIVE THAT DOCTOR CALLED AND STATED THAT WHEN HE DEPLOYED MAM3008 HE HAD A HARD TIME PULLING THE APPLICATOR OUT. WHEN HE DID PULL IT OUT IT LOOKED LIKE THE TIP WAS MISSING AND THERE WAS A PIECE OF COLLAGEN STUCK ON THE END. THEY X-RAY'D THE PATIENT AND IT LOOKS LIKE THE TIP IS INSIDE THE BREAST. THE REPRESENTATVIE LET THEM KNOW THAT IT APPEARS TO BE TIP SHEER. THE REPRESENTATIVE WILL REACH OUT TO THE DOCTOR TO FACILITY TO TRAIN ON HOW TO USE MAM3008. THIS INCIDENT IS RECORDED AS COMPLAINT NUMBER (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 15716214 ↗InjuryJan 7, 2022Use of Device Problem; Use of Device Problem
DURING A PROCEDURE A MARKER WAS DEPLOYED INTO THE PATIENT BUT IT DID NOT SHOW UP IN POST IMAGES SO THEY DEPLOYED A SECOND MARKER. THE SECOND MARKER WAS SUCCESSFULLY DEPLOYED BUT WHEN THE DOCTOR TRIED TO PULL THE PRODUCT OUT FROM THE BREAST THERE WAS RESISTANCE. AFTER THE PROCEDURE THEY COULD CONFIRM THAT THE DISTAL TIP OF THE APPLICATOR WAS STILL IN THE BREAST. THE DOCTOR IS REQUESTING INFORMATION TO SEE IF THE APPLICATOR TIP NEEDS TO BE REMOVED. THEY ARE WAITING ON THE RESULTS OF THE BIOPSY TO SEE IF THE PATIENT NEEDS A LUMPECTOMY AND THAT THEY COULD REMOVE THE APPLICATOR TIP AT THAT TIME AS WELL.
Patient problems: Foreign Body In Patient
MAUDE report 13197598 ↗MalfunctionFeb 6, 2020Mechanical Problem; Mechanical Problem
DEVICOR MEDICAL PRODUCTS, INC. RECEIVED A REPORT BY SALES REPRESENTATIVE OF A MAM3008 TIP SHEAR DURING A PROCEDURE. TIP OF DEPLOYMENT DEVICE CUT OFF IN PATIENT. THIS HAS BEEN DOCUMENTED IN OUR SYSTEM AS RECORD #(B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 9679771 ↗
Other Marker, Radiographic, Implantable devices
- BioZorb F0202Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls192 reports
- BioZorb F0221Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE3 recalls15 reports
- BioZorb F0203Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls51 reports
- BioZorb F0231Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls19 reports
- BioZorb F0303Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls19 reports
- BioZorb F0304Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls19 reports
- BioZorb F0404Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls6 reports
- BioZorb F0331Hologic, Inc. · MARKER, RADIOGRAPHIC, IMPLANTABLE2 recalls4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.