Skip to content

Hydromark CA000520005Recalls & reported problems

Devicor Medical Products, Inc. · HydroMark Breast Biopsy Site Marker -Non CE · Class 2 · Implant

All models of HydroMark (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841911103677), received between Aug 7, 2026 and Aug 7, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 7, 2026Use of Device Problem

    According to the reporter,  there were deployment issues with the HydroMark during a prone procedure using the Revolve 10g probe. The first marker did not deploy and confirmed that it was not stuck inside the applicator. Marker 1 Deployment Steps Confirmed by the Customer: 1. The aperture was rotated to 6:00, and the arrow was also pointing at 6:00. 2. The tech was not in the room, but she suspects that the marker applicator was not advanced until a definitive stop was felt. 3. The aperture was fully opened. 4. The plunger was pressed until a click was heard. 5. The probe and marker applicator were rotated 180 degrees away from deployment, and both were removed together.After this, another marker was inserted, which potentially led to a plunger shear. Following the procedure, a plastic fragment was discovered at the patient's biopsy site and was subsequently removed. Marker 2 Deployment Steps Confirmed by the Customer: 1. The aperture was rotated to 6:00, and the arrow was also pointing at 6:00. 2. The marker applicator was advanced until a definitive stop was felt. 3. The aperture was fully opened. 4. The plunger was pressed until a click was heard. 5. The probe and marker applica

    Patient problems: Foreign Body In Patient

    MAUDE report 50174848 ↗

Other Marker, Radiographic, Implantable devices

All Marker, Radiographic, Implantable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page