Neuromaster MEE-2000A-LTRecalls & reported problems
Nihon Kohden America, Inc. · MEE-2000 Neuromaster G1 Laptop system; performs Intraoperative and Evoked Potential studies · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00841983103629), received between Jul 3, 2019 and Apr 9, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Connection Problem3
- Failure to Deliver Energy2
- Loose or Intermittent Connection2
- Output Problem2
- Complete Loss of Power1
- Electrical /Electronic Property Problem1
- Energy Output Problem1
- Intermittent Shock/Stimulation1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2025Output Problem; Output Problem
THE CUSTOMER REPORTED THAT THE NEURO WORK BENCH WITH A MEE 2000A STIMULATOR FROZE DURING A PROCEDURE. THE CUSTOMER REPORTED THAT THE WIRELESS INTERNET DROPPED, WHICH CAUSED A VPN LOSS AND THIS MADE THE DEVICE(S) FREEZE. THE CUSTOMER REBOOTED THE DEVICE(S) AND AFTER THE REBOOT, THE NEURO WORK BENCH (NWB) INCLUDING MEE 2000A STIMULATOR SYSTEM WAS ABLE TO OPEN AND THEY WERE ABLE TO FINISH THE EXAM WITHOUT FURTHER ISSUES. NO PATIENT HARM WAS REPORTED. NOTE: THE MEE SYSTEM MEASURES AND DISPLAYS ELECTRIC, AUDITORY, AND VISUAL EVOKED POTENTIALS (EP), ELECTROENCEPHALOGRAMS (EEG), AND ELECTROMYOGRAPHS (EMG).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21810867 ↗MalfunctionFeb 7, 2024Failure to Deliver Energy; Energy Output Problem; Output Problem; Failure to Deliver Energy; Energy Output Problem; Output Problem
THE SURGICAL NURSE REPORTED THAT A PATIENT WAS UNDERGOING A SPINAL CORD STIMULATOR PLACEMENT PROCEDURE AND WHEN THE SURGICAL NURSE BEGAN MONITORING THE PATIENTS SOMATOSENSORY EVOKED POTENTIAL (SSEP'S) FOR DETECTION OF SPINAL CORD INJURY OR TETRALOGY OF FALLOUT (TOF) CARDIAC ANOMALY'S THERE WAS AN ERROR CODE THAT READ, "ELECTRICAL STIMULATOR TURNED OFF BECAUSE OF ERROR". RESTART THE ELECTRICAL STIMULATOR"?. THE SURGICAL NURSE PROCEEDED TO RESTART THE ELECTRICAL STIMULATOR AND STILL COULD NOT MONITOR THE PATIENT. THE SURGICAL NURSE REPORTED THAT THEY UNPLUGGED/PLUGGED THE STIM CABLE FROM THE STIM POD AS WELL AS THE MAIN UNIT. AFTER DOING THIS, THE DEVICE SHOWED ERROR CODES THAT SAID "YOU CANNOT RESTART THE AUTOMATIC TREND SEQUENCE, BECAUSE ELECTRIC A1-A4 ARE NOT CONNECTED. PLEASE CHECK THE STIMULATOR IS CONNECTED TO THE CONNECTOR ON THE RIGHT SIDE OF THE MAIN UNIT". THE SURGICAL NURSE NOTED THAT THE LIGHT ON THE STIM POD WAS NOT ON. THE SURGEON WAS AWARE THAT THE SURGICAL NURSE WAS HAVING TECHNICAL ISSUES DURING THE PROCEDURE. THE SURGICAL NURSE STATED THAT SINCE THE PROCEDURE WAS SO QUICK, THEY WERE NOT ABLE TO GET SPARE EQUIPMENT IN TIME, SO THEY WERE UNABLE TO MONITOR THE SOMATOSE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18663606 ↗MalfunctionJan 17, 2024Loose or Intermittent Connection; Connection Problem; Loose or Intermittent Connection; Connection Problem
THE CUSTOMER REPORTED THAT THE STIMULATION POD WAS NOT STAYING CONNECTED DURING A PATIENT'S EXAMINATION AND WOULD NOT RECONNECT. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18531524 ↗MalfunctionJan 10, 2024Loose or Intermittent Connection; Connection Problem; Loose or Intermittent Connection; Connection Problem
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE MEE-2000A MAIN UNIT HAD A BAD AMPLIFIER CABLE. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18485659 ↗MalfunctionJan 6, 2024Connection Problem; Unexpected Shutdown; Connection Problem; Unexpected Shutdown
THE BIOMEDICAL ENGINEER (BME) REPORTED HAVING ISSUES WITH THE PATCH CABLE CONNECTION BETWEEN THEIR LAPTOP AND THE NK MAIN UNIT. THEY RECEIVED AN ERROR MESSAGE STATING CONNECTION ERROR FAILURE WHEN STARTING THE MAIN UNIT. HOWEVER, IT HAPPENED TWO TIMES DURING A PROCEDURE, AND THEY HAD TO REBOOT AND HOLD THE CABLE IN TO GET IT TO START. THEY TRIED SWITCHING BETWEEN MULTIPLE CABLES, BUT THE ERROR STILL OCCURRED. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18461132 ↗MalfunctionJul 17, 2019Electrical /Electronic Property Problem; Failure to Deliver Energy; Intermittent Shock/Stimulation; Electrical /Electronic Property Problem; Failure to Deliver Energy; Intermittent Shock/Stimulation
THE EEG TECH REPORTED THAT THE JB-201B STIMULATION POD WAS DELIVERING INTERMITTENT STIMULUS WHILE CONNECTED TO THE MEE-2000 DURING A PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 8800363 ↗MalfunctionJul 3, 2019Power Problem; Complete Loss of Power; Power Problem; Complete Loss of Power
THE EEG TECH REPORTED THAT THE MEE SYSTEM SHUT DOWN AND RESTARTED DURING A 4EXT CERVICAL PROCEDURE. THE MEE-2000A IS AN IOM SYSTEM WHICH IS USED FOR EVOKED POTENTIALS, EEGS, AND EMGS.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 8761452 ↗
Other STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices
- NIM Vital™ NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls795 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls331 reports
- NIM Vital NIM4CPB1Medtronic Xomed, Inc. · Stimulator, electrical, evoked response2 recalls208 reports
- POWEREASE™ Instruments 2300000Medtronic Sofamor Danek, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE1 recall203 reports
- NIM Vital™ NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls47 reports
- NIM-Response® 8253200Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE113 reports
- NA DAQ916Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE80 reports
- Neuromaster G1 DC-200BNihon Kohden Corporation · STIMULATOR, ELECTRICAL, EVOKED RESPONSE73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page