Skip to content

Cyclops Anterior Cervical Plate System 0220-4068TRecalls & reported problems

Degen Medical, Inc. · 4-Level Plate, Size 68 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842048121817), received between Nov 13, 2020 and Nov 13, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2020Break; Break

    DEGEN MEDICAL WAS NOTIFIED THAT A CERVICAL SCREW BACKED OUT OF A 4-LEVEL CYCLOPS ANTERIOR CERVICAL SYSTEM PLATE. THE ORIGINAL SURGERY OCCURRED ON (B)(6) 2020. RADIOGRAPHIC IMAGES DURING A POST-OPERATION FOLLOW-UP WITH THE PATIENT SHOWED THAT ONE CERVICAL SCREW BACKED OUT FROM THE PLATE. A REVISION SURGERY WAS COMPLETED ON (B)(6) 2020 TO REMOVE THE EXISTING PLATE AND SCREWS AND TO IMPLANT A NEW CERVICAL PLATE, AND SCREWS. DURING THE REMOVAL OF THE ORIGINAL HARDWARE AS PART OF THE REVISION SURGERY IT WAS FOUND THAT THE LOCKING MECHANISM, THAT WAS INTENDED TO COVER THE BACKED OUT SCREW, WAS IN THE OPEN POSITION. THE LOCKING MECHANISM HAD BROKEN AWAY FROM THE PLATE. NO PATIENT HARM WAS REPORTED FROM THIS EVENT. THIS REPORT CAPTURES THE PLATE. AN ADDITIONAL REPORT IS CREATED FOR THE SCREW.

    Patient problems: Failure of Implant

    MAUDE report 10836221 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page