Skip to content

Traditional G-Jet® TGJ-18-45Recalls & reported problems

Applied Medical Technology, Inc. · 18FR x 45 cm - AMT Traditional G-Jet® Transgastric-Jejunal Feeding Device · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842071125318), received between Mar 31, 2021 and Mar 31, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Remove1
  • Fracture1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2021Fracture; Difficult to Remove

    IMAGING OVER THE ABDOMEN DEMONSTRATED APPARENT FRACTURING OF THE GJ TUBE. CONTRAST WAS INJECTED THROUGH THE TUBE DEMONSTRATING COMPLETE FRACTURING OF THE GJ TUBE NEAR THE GASTRIC LUMEN WITH COMPLETE DISCONTINUITY OF THE JEJUNAL PORTION OF THE CATHETER. A.035 STIFF GLIDEWIRE WAS ADVANCED THROUGH THE CATHETER INTO THE STOMACH. THE GASTRIC PORTION OF THE TUBE WAS REMOVED. THE DISTAL JEJENAL PORTION WAS RETAINED INSIDE THE PATIENT AND WAS CONNECTED TO THE PROXIMAL GASTRIC PORTION VIA A METALLIC SPRING COIL. ATTEMPT WAS MADE TO PULL THE DISTAL PORTION OUT VIA THE COIL, HOWEVER THIS CAUSE SIGNIFICANT PAIN FOR THE PATIENT. AFTER FURTHER DISCUSSION, DR. OF GI WAS CONSULTED. THE DECISION WAS MADE TO HAVE THE RETAINED PORTION OF THE CATHETER REMOVED ENDOSCOPICALLY BY GI. THE METALLIC SPRING COIL WAS CUT WITH SCISSORS, AND THE EXTERNAL PORTION TAPED TO THE PATIENT'S SKIN.

    Patient problems: Pain

    MAUDE report 11598945 ↗

Other Tubes, Gastrointestinal (And Accessories) devices

All Tubes, Gastrointestinal (And Accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page