Skip to content

G-Jet® Transgastric-Jejunal Feeding Device GJ-1430-45-IRecalls & reported problems

Applied Medical Technology, Inc. · 14FR x 3.0 cm x 45 cm Low Profile Transgastric-Jejunal Feeding Device (Enfit), MDR · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842071134310), received between Feb 20, 2026 and Feb 20, 2026 — classified as 1 malfunction.

1 other recall cover devices cleared under the same FDA submission (K110804, K123716); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026No Apparent Adverse Event

    PER THE ORIGINAL REPORTER IN MEDSUN REPORT #: (B)(4), IT WAS REPORTED THAT THE, "RN WENT TO UNCLOG THE G-TUBE PORTION OF THE GJ [GASTRO-JEJUNAL] TUBE BY INSTILLING WARM WATER INTO THE EXTENSION, SINCE THE PATIENT COULD NOT RECEIVE CLOG ZAPPER DUE TO THERE BEING MALTODEXTRIN IN THE ZAPPER INSTILLATION (PT ALLERGY). THE WARM WATER COULD NOT BE ADVANCED INTO THE TUBE, SO THIS RN IRRIGATED THE G-PORT WITH SOME STERILE WATER AND GAUZE TO WASH AWAY SOME CLOGGED RESIDUE, AS IT COULD BE SEEN AT THE OPENING TO THE G-TUBE PORT. AFTER IRRIGATING SOME AWAY, THIS RN REATTACHED THE G-TUBE EXTENSION AND FLUSHED WITH WARM STERILE WATER. AS IT WAS BEING FLUSHED, THE TUBE STARTED LEAKING UNDERNEATH THE MIC BUTTON, WHERE THE G-TUBE PORTION OF THE GJ GOES INTO THE PATIENT. IT WAS DISCOVERED THAT A HOLE WAS UNDERNEATH THE MIC-BUTTON. PT UNDERWENT GJ TUBE EXCHANGE OVER WIRE UNDER SEDATION".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403232 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K110804, K123716. One submission often covers many models, so check the product and lot details in each record.

  • Z-1497-2016 · Jan 26, 2016 · Terminated
    AMT G-Jet Low-Profile Transgastric-Jejunal Feeding Device 14F 1.5cm 45cm SINGLE USE ONLY Rx ONLY STERILE. For use in patients who cannot absorb adequate nutrition through the stoma

Other Tubes, Gastrointestinal (And Accessories) devices

All Tubes, Gastrointestinal (And Accessories) devices →

Sources and updates