Skip to content

NxStage PureFlow-B Solution RFP-400-BBRecalls & reported problems

Nxstage Medical, Inc. · Class 2

All models of NxStage PureFlow-B Solution (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Nov 10, 2017Terminated

    Using different fluid formulations of NxStage PureFlow B Solution at the same time during treatment may cause patient health risk. The NxStage PureFlow B Solution is for use with systems that use one premixed dialysate during hemodialysis. The original labeling did not warn against mixing dialysate formulations and did not instruct staff to verify that the fluid is flowing during use.

    Z-1132-2018 · Root cause: Labeling design · Recall details · FDA record ↗

Other Dialysate Concentrate For Hemodialysis (Liquid Or Powder) devices

All Dialysate Concentrate For Hemodialysis (Liquid Or Powder) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page