Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842429101582), received between Oct 16, 2020 and Sep 9, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Complete Blockage1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2024Complete Blockage; Complete Blockage

    WHILE PREPARING THE SOFIA DEVICE ON THE BACK TABLE, THE MICROCATHETER WOULD NOT ADVANCE THROUGH THE SOFIA EX. THE OPERATOR THEN ATTEMPTED TO ADVANCE JUST A WIRE, BUT THAT ALSO WOULD NOT ADVANCE. THE SOFIA WAS NOT USED AND WAS REPLACED WITH A COMPETITOR DEVICE. ADDITIONAL INFORMATION FROM THE POST-EVALUATION REPORT REVEALED THAT A PTFE LINER WAS FOUND TO BE OCCLUDING THE LUMEN, WHICH IS IN LINE WITH THE DESCRIBED ADVANCEMENT ISSUE IN THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20177479 ↗
  • MalfunctionApr 27, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT DURING THE TREATMENT OF AN ANEURYSM OF THE ANTERIOR COMMUNICATING ARTERY, THE ANATOMY WAS VERY TORTUOUS AND IT WAS DECIDED TO EXCHANGE A SOFIA EX CATHETER FOR A LONGER ONE. WHILE REMOVING THE SOFIA EX OVER THE EXCHANGE WIRE, THE PHYSICIAN NOTICED THE TIP OF THE SOFIA EX WAS DAMAGED AND MISSING ITS DISTAL MARKER BAND. AS HE MANIPULATED THE WIRE TO INTRODUCE THE NEW CATHETER, THEY NOTICED THE DISTAL MARKER WAS OUTSIDE OF THE SHEATH AND WRAPPED AROUND THE EXCHANGE WIRE. THERE WAS NO INTERVENTION OR PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11735558 ↗
  • InjuryOct 16, 2020Break; Break

    IT WAS REPORTED THAT DURING USE IN THE PATIENT, THE TIP OF THE CATHETER DETACHED DURING RETRACTION INTO THE INFINITY SHEATH. AN ATTEMPT WAS MADE TO RETRIEVE THE TIP WITH A STENT RETRIEVER, BUT IT WAS UNSUCCESSFUL. THE TIP REMAINS IN THE PATIENT. THERE WAS NO REPORTED SEQUELAE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10694063 ↗

Other Catheter, Percutaneous devices

All Catheter, Percutaneous devices →