Balloon Catheter BC0415C-ID-700Recalls & reported problems
Microvention, Inc. · Scepter C · Class 2
All models of Balloon Catheter (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00842429103395), received between Jan 13, 2021 and Jul 3, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Deflation Problem2
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2024Deflation Problem; Deflation Problem
IT WAS REPORTED THAT A SCEPTER BALLOON CATHETER DID NOT DEFLATE DURING A COIL EMBOLIZATION PROCEDURE. THE DEVICE WAS REMOVED IN AN INFLATED STATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE AND THE PATIENT WAS REPORTED TO BE STABLE. THERE WAS NO INJURY OR INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19662926 ↗MalfunctionMay 9, 2024Material Rupture; Material Rupture
AS REPORTED: DURING THE BALLOON-ASSISTED-COILING OF A MEDIA ANEURYSM, THE BALLOON BURST WITHIN THE VESSEL. THE BALLOON WASN'T INFLATED COMPLETELY ¿ AROUND 50-70%. WHEN THE BALLOON HAD BURST, THERE WAS A QUICK LOSS OF VOLUME IN A SHORT TIME. THE EVENT HAD NO INFLUENCE ON THE PATIENT'S OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19280646 ↗MalfunctionApr 1, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING ATTEMPT TO INFLATE THE SCEPTER BALLOON INSIDE TARGET VESSEL (ICA) UNDER FLUOROSCOPY, THE BALLOON DID NOT APPEAR TO INFLATE AND WAS REMOVED FROM THE PATIENT, WHERE IT WAS FOUND TO HAVE RUPTURED. THERE WAS NO REPORT OF HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13983878 ↗MalfunctionMar 26, 2022Fluid/Blood Leak; Material Rupture; Fluid/Blood Leak; Material Rupture
IT WAS REPORTED THAT THE SCEPTER BALLOON EXHIBITED A RAPID PRESSURE LOSS AND CONTRAST LEAKAGE DURING AN OCCLUSION TEST PRIOR TO A CAROTID OCCLUSION. THERE WAS NO PATIENT HARM REPORTED. THE PATIENT IS RELEASED FROM THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13923624 ↗MalfunctionAug 9, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING BALLOON INFLATION WITHIN A FLOW DIVERTER STENT, THE BALLOON RAPIDLY LOST PRESSURE. THE BALLOON CATHETER WAS REMOVED IN ITS ENTIRETY FROM THE PATIENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12298386 ↗MalfunctionJan 13, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED THAT THE BALLOON DID NOT CORRECTLY DEFLATE DURING THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY OR ADDITIONAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11168473 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page