VIA Microcatheter VIA-17-154-01-MVI2Recalls & reported problems
Microvention, Inc. · VIA 17 Microcatheter · Class 2
All models of VIA Microcatheter (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00842429106402), received between Sep 23, 2021 and Apr 24, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Unintended Movement2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Ruptured Aneurysm4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryMay 27, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING TREATMENT OF THE ACOMM ANEURYSM THE VIA-17 CATHETER RE-RUPTURED THE ANEURYSM. THE PHYSICIAN PLACED A SCEPTER XC AND A WEB DEVICE TO TREAT THE PATIENT. THE STATUS OF THE PATIENT IS UNKNOWN AT THIS TIME.
Patient problems: Ruptured Aneurysm
MAUDE report 14526091 ↗InjuryMar 23, 2022Unintended Movement; Unintended Movement
IT WAS REPORTED THAT DURING TREATMENT OF ANEURYSM ON THE LEFT ICA (CAROTID T), THE IMPLANTATION OF A WEB DEVICE WAS PLANNED. DURING THE ATTEMPT TO CANNULATE THE ANEURYSM, THE CONTROL OVER THE VIA-17 MICROCATHETER WAS LOST AND THE DEVICE MOVED FORWARD AND PERFORATED THE ANEURYSM DUE TO A STEEP ANGLE FROM THE PARENT VESSEL TO THE ANEURYSM. IT WAS DECIDED TO USE A WEB DEVICE FOR A QUICK TREATMENT OF THE BLEEDING. THE FIRST WEB DEVICE DID NOT FIT TO THE ANEURYSM AND WAS REMOVED. THE WEB WAS REPLACED BUT THE WEB AND VIA WERE REMOVED FROM THE PATIENT DUE TO PLACEMENT DIFFICULTY. THE ANEURYSM WAS COILED SUCCESSFULLY. THE COILS DEPLOYED PARTIALLY THROUGH THE PERFORATION. THE HEMORRHAGE WAS STOPPED, BUT THE ANEURYSM WAS NOT TREATED SUCCESSFULLY. THE PATIENT DID HAVE AN EVD BECAUSE OF THE HEMORRHAGE TWO WEEKS PRIOR. FOR THIS REASON THE ANESTHESIOLOGIST WAS ABLE TO DRAIN A CERTAIN AMOUNT OF BLOOD FROM THE BRAIN. AFTER THE EXAMINATION, A CT WAS PERFORMED REVEALING SWELLING OF BOTH VENTRICLES. THE NEUROSURGEON DECIDED AGAINST AN OPERATION TO PLACE A SECOND EVD, BECAUSE THE PATIENT WAS STABLE AND THE HEMORRHAGE STOPPED.
Patient problems: Ruptured Aneurysm
MAUDE report 13865100 ↗InjurySep 23, 2021Unintended Movement; Unintended Movement
IT WAS REPORTED THAT A PATIENT WITH A RUPTURED MCA ANEURYSM WAS TREATED WITH A WEB DEVICE. THE PHYSICIAN PLACED A VIA 17 MICROCATHETER IN THE MIDDLE OF THE ANEURYSM AND REMOVED THE GUIDEWIRE. THE PHYSICIAN THEN ADVANCED THE WEB THROUGH THE VIA 17 UNDER FLUOROSCOPY. DURING ADVANCEMENT, THE VIA 17 MOVED AND WENT THROUGH THE DOME OF THE ANEURYSM AT THE RUPTURE SITE. THE PHYSICIAN THEN PLACED THE WEB DEVICE IN THE ANEURYSM. THE PATIENT IS REPORTED TO BE "DOING VERY WELL AND WILL BE DISCHARGED SOON." THERE WAS NO REPORTED INTERVENTION.
Patient problems: Ruptured Aneurysm
MAUDE report 12523232 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page