VIA Microcatheter VIA-27-154-01-MVI2Recalls & reported problems
Microvention, Inc. · VIA 27 Microcatheter · Class 2
All models of VIA Microcatheter (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842429106426), received between Jun 18, 2021 and May 30, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Peeled/Delaminated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Peeled/Delaminated1
- Unintended Movement1
Patient problems
- Extravasation1
- Headache1
- Hemorrhage/Blood Loss/Bleeding1
- No Clinical Signs, Symptoms or Conditions1
- Ruptured Aneurysm1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 30, 2025Peeled/Delaminated; Peeled/Delaminated
ADDITIONAL INFORMATION WAS RECEIVED THAT STATED IT HAPPENED WHILE INSERTING THE VIA INTO THE HUB. AFTER THAT, THEY TOOK ANOTHER VIA AND USED IT WITH THE WEB TO COMPLETE THE CASE. THERE WERE NO PICTURES TAKEN OF THE SEGMENT THAT WAS IN THE CATHETER HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22118025 ↗InjuryOct 6, 2022Unintended Movement; Unintended Movement
IT WAS REPORTED THAT WHEN ADVANCING A WEB DEVICE THROUGH THE VIA 27 DURING THE TREATMENT OF AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM, THE VIA CATHETER SHIFTED FORWARD AND PERFORATED THE DOME OF THE ANEURYSM, RESULTING IN BLEEDING INTO THE BRAIN. THE VIA AND WEB DEVICE WERE REMOVED, AND THE RUPTURED ANEURYSM WAS TREATED WITH COILS. THE PATIENT WAS BEING MONITORED IN THE ICU AND REPORTED TO BE NEUROLOGICALLY INTACT AND HAD A HEADACHE. THERE WERE NO FOCAL DEFICITS OR HYDROCEPHALUS, AND THE BLOOD IS CLEARING UP ON CT HEAD.
Patient problems: Headache; Hemorrhage/Blood Loss/Bleeding; Ruptured Aneurysm
MAUDE report 15556223 ↗InjuryJun 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING TREATMENT OF AN ANEURYSM, A VIA MICROCATHETER AND WEB DEVICE WERE ADVANCED INTO POSITION WHEN AN ANGIOGRAPHIC RUN REVEALED EXTRAVASATION. IT WAS NOT KNOWN IF THE WEB OR THE VIA CAUSED THE EXTRAVASATION. HEPARIN WAS REVERSED AND THE NEXT ANGIOGRAPHIC RUN REVEALED NO FURTHER EXTRAVASATION. THE WEB WAS IMPEDING FLOW IN THE PARENT VESSEL. A COMPETITOR'S STENT WAS PLACED IN ORDER TO KEEP THE WEB DEVICE IN THE ANEURYSM. FLOW THROUGH THE STENT SHUT DOWN AND INTEGRILIN WAS ADMINISTERED, WHICH RESTORED FLOW. POST PROCEDURE, THE PATIENT WAS AWAKE, FOLLOWING COMMANDS AND MOVING THEIR RIGHT AND LEFT LEGS. THERE WAS NO REPORTED SEQUELAE.
Patient problems: Extravasation
MAUDE report 12029483 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page