Stent FRED3021-PMARecalls & reported problems
Microvention, Inc. · FRED 21 · Class 3 · Implant
All models of Stent (63)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00842429106631), received between Jul 24, 2020 and May 26, 2021 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Activation Failure1
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Extravasation1
- No Consequences Or Impact To Patient1
- Speech Disorder1
- Thrombosis/Thrombus1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMay 26, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT THE FRED FLOW DIVERTER STENT WAS USED IN THE PATIENT TO TREAT PREVIOUSLY DEPLOYED EMBOLIZATION DEVICE FROM A DIFFERENT MANUFACTURER THAT MIGRATED AND RESULTED IN AN ENDOLEAK. THE FRED WAS PLACED THROUGH THE EMPOLIZATION DEVICE FROM A2 TO A1 SEGMENT. ANGIOGRAM RUN IMMEDIATELY AFTER DEPLOYMENT LOOKED GOOD, BUT LATER RUNS SHOWED SLOW FLOW BACK THROUGH THE VESSEL. DISTAL END OF FRED LOOKED NARROW UPON IMAGING REVIEW. CROSSING THE FRED WAS ATTEMPTED WITH A WIRE; HOWEVER, A MICROCATHETER COULD NOT INITIALLY CROSS. EVENTUALLY A BALLOON CATHETER WAS SUCCESSFULLY ADVANCED INTO THE FRED AND INFLATED IN THE DISTAL END RESULTING IN RESTORATION OF NORMAL FLOW. THE PATIENT WAS STABLE AND AWAKE POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11891040 ↗InjuryDec 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT 12 HOURS AFTER IMPLANTATION OF THE FRED IN THE SUPRACLINOID SEGMENT OF THE ICA, THROMBUS WAS IDENTIFIED IN THE ARTERY. THE PATIENT EXPERIENCED DYSARTHRIA. REOPRO AND HEPARIN WERE ADMINISTERED TO THE PATIENT. THERE WAS NO REPORTED MALFUNCTION OF THE FRED.
Patient problems: Speech Disorder; Thrombosis/Thrombus
MAUDE report 11042424 ↗InjuryOct 23, 2020Unintended Movement; Unintended Movement
IT WAS REPORTED THAT DURING IMPLANTATION OF THE FRED JR, THE DISTAL END OF THE STENT PULLED BACK "INTO THE DEPLOYMENT." CONTRAST EXTRAVASATION WAS OBSERVED. THE FRED WAS RESHEATHED AND REMOVED, HEPARIN WAS REVERSED, AND THE ANEURYSM WAS COILED. THE EXTRAVASATION WAS RESOLVED. AN EVD WAS PLACED. FOLLOW-UP CT SHOWED THE BRAIN WAS "NORMAL," HOWEVER, A TRACT HEMORRHAGE WAS NOTED FROM THE EVD PLACEMENT.
Patient problems: Extravasation
MAUDE report 10731620 ↗MalfunctionJul 24, 2020Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT TREATMENT WAS PERFORMED FOR AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM. THE ORIGIN OF THE ACOM COMPLEX WAS REPORTED TO BE SLIGHTLY TORTUOUS WITH SOME NARROWING NEAR THE NECK OF THE ANEURYSM. AFTER PLACEMENT OF THE FRED BETWEEN THE A2 AND A1 SEGMENT, THE PUSHER WIRE CAUGHT ON THE STENT DURING THE REMOVAL ATTEMPT. THE INTERMEDIATE CATHETER WAS ADVANCED AND USED TO DETACH THE STENT FROM THE PUSHER WIRE AND MICROCATHETER. AFTER DEPLOYMENT, THE FRED SLIGHTLY EXTENDED INTO THE INTERNAL CAROTID ARTERY. THE STENT REMAINS IMPLANTED. NO ADDITIONAL INTERVENTION WAS PERFORMED AND THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10320702 ↗
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.