HEADWAY MC172150SX-ALRecalls & reported problems
Microvention, Inc. · Headway Advanced · Class 2
All models of HEADWAY (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842429117439), received between Apr 24, 2025 and Sep 23, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Break
It was reported that when the microcatheter Headway-17 was opened the device was straight, but when the microguide was advanced by 0.014 mm, the doctor realized it wasn't progressing. He tried repeatedly, but it was never possible. The device was inspected and found to be fractured in the proximal portion. Another device is used to resolve the case. The patient is in good clinical condition.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872020 ↗MalfunctionDec 4, 2025Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT THE GUIDE WIRE DID NOT PASS SMOOTHLY THROUGH THE MICROCATHETER. A NEW CATHETER WAS USED TO COMPLETE THE CASE. THE PATIENT IS IN STABLE CONDITION. ADDITIONAL INFORMATION WAS RECEIVED FROM THE POST EVALUATION REPORT REVEALED AN EXPOSED COIL AND DAMAGED/MISSING PTFE LINING AT THE HUB TRANSITION, WHICH WOULD HAVE CONTRIBUTED TO THE RESISTANCE ENCOUNTERED DURING THE INVESTIGATION. THE PTFE DAMAGE/DELAMINATION AND RESULTING OCCLUSION ARE CONSISTENT WITH ADVANCING ANOTHER DEVICE THROUGH THE MICROCATHETER WITH EXPOSED COIL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23717878 ↗InjuryApr 24, 2025Break; Break
ADDITIONAL CLARIFYING INFORMATION RECEIVED NOTES THAT THE TYPE OF CASE PERFORMED WAS FOR A STENT COIL PROCEDURE. THE HEADWAY MICROCATHETER WAS PLACED WITHIN THE ANEURYSM, AND THE STENT WAS PLACED USING ANOTHER HEADWAY MICROCATHETER. AFTER COMPLETING THE COILING, THE DOCTOR TRIED TO REMOVE THE MICROCATHETER, BUT IT CAUGHT THE STENT CELL AND THE MICROCATHETER DISTAL TIP WAS TORN OFF. THE PROCEDURE WAS FINISHED WITH THE DISTAL TIP OF THE MICROCATHETER REMAINING IN THE PATIENT BODY. THERE IS NO PLAN TO ADDRESS PIECE REMAINING IN THE PATIENT AND WE WILL WATCH THE PROGRESS OF THE PATIENT. THE DEVICE IS CURRENTLY PENDING RECEIPT.
Patient problems: Foreign Body In Patient
MAUDE report 21906174 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.