SCEPTER BC0411XC-ID-700-ALRecalls & reported problems

Microvention, Inc. · Scepter XC · Class 2

All models of SCEPTER (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00842429117583), received between Apr 30, 2024 and Nov 7, 2025 — classified as 1 death, 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture4
  • Deflation Problem2
  • Failure to Deflate2
  • Adverse Event Without Identified Device or Use Problem1
  • Contamination1
  • Failure to Deliver1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2025Material Rupture; Failure to Deflate; Material Rupture; Failure to Deflate

    AS REPORTED VIA MEDWATCH # MW5176421: RUPTURED BALLOON ON CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23503971 ↗
  • DeathNov 3, 2025Deflation Problem; Deflation Problem

    IT WAS REPORTED TO MICROVENTION THAT THE PHYSICIAN DID NOT THINK THE BALLOON FULLY DEFLATED, WHICH RESULTED IN THE PHYSICIAN FINDING IT CHALLENGING TO TRACK THE CATHETER TO THE OPTIMAL LOCATION. REPORTEDLY THE DEVICE DID NOT PUNCTURE ANY ORGAN OR TISSUE, HOWEVER IT WAS THOUGHT THE DEVICE ¿DID NOT GET TO WHERE IT NEEDED TO BE¿ BECAUSE OF THE REPORTED DEFLATION CONDITION. THERE WAS NO INTERVENTION NOTED FOR THE SURGERY PERFORMED (B)(6) 2025, HOWEVER IT IS REPORTED THE PATIENT PASSED AWAY WITHIN 24 HOURS OF THE SURGERY. REPORTEDLY, THE CAUSE OF DEATH WAS SUBARACHNOID HEMORRHAGE. NO FURTHER INFORMATION WAS PROVIDED. THE DEVICE WAS REPORTEDLY DISCARDED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 23458436 ↗
  • InjuryMar 26, 2025Failure to Deliver; Adverse Event Without Identified Device or Use Problem; Failure to Deliver; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A CEREBRAL EMBOLIZATION, THE DOCTOR HAD THE COMPLAINT BALLOON PREPARED UNDER IFU. HOWEVER, UPON ATTEMPTING TO INJECT ONYX THROUGH THE BALLOON, IT WAS NOT ABLE TO INJECT ANY ONYX AT ALL. EMBOLIZATION WAS ABORTED AS THE RESULT OF THAT. IT IS UNKNOWN WHY ANOTHER DEVICE WAS NOT USED TO COMPLETE THE CASE. THE PATIENT CURRENT CONDITION WAS REPORTED AS LEFT UNTREATED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21704294 ↗
  • MalfunctionDec 11, 2024Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THAT: AFTER USING THE BALLOON AT THE NECK OF THE OF THE SYLVIAN ANEURYSM TO BE EMBOLIZED, THE PHYSICIAN FOUND IT WAS IMPOSSIBLE TO DEFLATE THE BALLOON. THE SCEPTER WAS REMOVED GENTLY AND VERY SLOWLY TO AVOID ANY DAMAGE TO THE ARTERY WALL. THE PHYSICIAN USED ANOTHER SCEPTER TO PERFORM / COMPLETE THE PROCEDURE. NO ACTUAL PATIENT IMPACT OR INJURY IS DOCUMENTED, AND IT WAS CONFIRMED PATIENT OUTCOME WAS GOOD AT THE END OF THE PROCEDURE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20909099 ↗
  • MalfunctionNov 8, 2024Material Rupture; Material Rupture

    AS REPORTED: PATIENT UNDERGOING FURTHER EMBOLIZATION OF VERTEBROBASILAR ANEURYSM. AFTER PREPARATION OF THE BALLOON IN ACCORDANCE WITH RECOMMENDATIONS, THE BALLOON WAS FOUND TO BE POROUS AND STRIATED, AND AIR ENTRY WAS VISUALIZED WHILE BEING USED INSIDE THE PATIENT. THE PATIENT IS WELL AFTER THE PROCEDURE. ADDITIONAL INFORMATION RECEIVED: THERE WAS NO AIR INSIDE THE BALLOON, THIS ONE WAS A LITTLE BIT STRETCHED WHEN THEY PULL IT BACK AFTER IT BROKE INSIDE THE PATIENT WITH PRODUCT OF CONTRAST AND SERUM PHYSIOLOGIC. WHEN THE BALLOON BROKE THE PHYSICIAN JUST PULL IT BACK OF THE PATIENT WITHOUT DAMAGE. THE PATIENT WENT WELL AND THEY PERFORMED THE PROCEDURE WITH ANOTHER BALLOON OF THE COMPETITORS. THE PHYSICIAN IS COMPLAINING ABOUT THE BALLON BREAKAGE WITHOUT AIR. THERE IS NO IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20637512 ↗
  • MalfunctionOct 29, 2024Contamination; Contamination

    SEE H10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20559527 ↗
  • MalfunctionOct 21, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PHYSICIAN WERE TREATING AN ANEURYSM ON THE RIGHT ICA. A COIL EMBOLIZATION IN REMODELING TECHNIQUE WAS PLANNED. A SCEPTER XC 4-11 BALLOON WAS CHOSEN FOR THE PROCEDURE. THE BALLOON WAS PREPARED AND THE TIP WAS SEALED BY USING THE HEAT GUN. THE BALLOON WAS PLACED AT THE NECK OF THE ANEURYSM. A COILING MICROCATHETER WAS PLACED INSIDE THE ANEURYSM. AFTER PLACING THE TWO COILS THE MICROCATHETER LOST ITS POSITION. THE BALLOON WAS DEFLATED AND THE COILING MICROCATHETER REPOSITIONED. WHEN THE PHYSICIAN INFLATED THE BALLOON AGAIN IT WAS ALMOST NOT VISIBLE. SHE REMOVED THE BALLOON FROM THE PATIENT AND CHECKED IT. SHE FOUND BLOOD INSIDE THE BALLOON. THE TIP WAS SEALED WITH THE HEAT GUN FOR A SECOND TIME. THE BALLOON WAS PLACED AT THE NECK OF THE ANEURYSM. DURING THE INFLATION THE DISTAL TIP OF THE BALLOON BURST. THE BALLOON WAS EXCHANGED TO A NEW PRODUCT. THE EXAMINATION WAS CONTINUED WITH GOOD RESULT. THE PATIENT IS DOING WELL. NO INJURY CAUSED BY THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20501113 ↗
  • MalfunctionSep 28, 2024Deflation Problem; Deflation Problem

    AS REPORTED, DURING AN M1 ANEURYSM EMBOLIZATION FOLLOWING SUBARACHNOID HEMORRHAGE, THE PHYSICIAN COULDN¿T DEFLATE THE COMPLAINT SCEPTER XC BALLOON. THE DEVICES USED WERE: BENCHMARK, HEADWAY 17, SYNCHRO STANDARD WIRE, ASAHI CHIKAI WIRE. THE COMPLAINT SCEPTER XC BALLOON WAS PREPPED UNDER IFU. THE CONTRAST AND SALINE RATIO IN BALLOON WAS 60% TO 40%, OMNIPAQUE 240 WAS USED. THE PHYSICIAN IS AN EXPERIENCE SCEPTER USER. AFTER POSITIONING THE COMPLAINT SCEPTER XC BY THE SIDE OF THE M1 ANEURYSM, HE INFLATED THE BALLOON TO TEST IT BEFORE HE COMMENCED COILING. HOWEVER, THE BALLOON FAILED TO BE DEFLATED. ATTEMPTS WERE MADE TO DEFLATE THE BALLOON WITH 1,5, AND 20ML SYRINGES. THE PHYSICIAN HAD TO REMOVE THE BALLOON WHILE IT WAS INFLATED FROM M1. THE EMBOLIZATION WAS ACHIEVED WITH ANOTHER SCEPTER XC BALLOON AND COILS. NO IMMEDIATE ADVERSE EVENT HAPPENED ON PATIENT DUE TO THIS COMPLAINT PRODUCT. THE CASE WAS COMPLETED WITH ANOTHER SCEPTER XC BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20328833 ↗
  • MalfunctionApr 30, 2024Material Rupture; Material Rupture

    A CORRECTION HAS BEEN MADE TO SECTION D2A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19217752 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →