VIA Microcatheter VIA-17-154-01-ALRecalls & reported problems
Microvention, Inc. · VIA 17 Microcatheter · Class 2
All models of VIA Microcatheter (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842429118702), received between Jul 14, 2025 and Jul 23, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement2
Patient problems
- Intracranial Hemorrhage2
- Aneurysm1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 23, 2025Unintended Movement; Unintended Movement
IT WAS REPORTED THAT A MEDIA ANEURYSM TREATMENT WAS PERFORMED FOR A PATIENT THAT HAS A VERY ELONGATED/TORTUOUS VESSEL ANATOMY. THE CATHETER, TOGETHER WITH THE GUIDEWIRE, WAS POSITIONED INTO THE ANGLED ANEURYSM SUCCESSFULLY. AFTER PULLING SLIGHTLY THE GUIDEWIRE OUT OF THE CATHETER, THE CATHETER MOVED FORWARD AND PERFORATED THE ANEURYSM, AND AN SAH OCCURRED. AN EMERGENCY COILING WAS PERFORMED TO STOP THE BLEEDING. AFTER THAT THE PATIENT UNDERWENT A NEUROSURGERY FOR EXTERNAL VENTRICULAR DRAIN (EVD). THE PATIENT¿S CONDITION IS UNKNOWN. IMAGE AND OPERATIVE REPORT WERE REQUESTED, AND NO RESPONSE WAS RECEIVED.
Patient problems: Intracranial Hemorrhage
MAUDE report 22576052 ↗InjuryJul 14, 2025Unintended Movement; Unintended Movement
IT WAS REPORTED THAT WHILE MOVING THE WEB DEVICE IN THE VIA MICROCATHETER, IT WAS NOTED UNDER FLUOROSCOPY THAT THE VIA HAD MOVED THROUGH THE ANEURYSM AND THE PATIENT WAS HEMORRHAGING. THE WEB WAS IMMEDIATELY REMOVED FROM THE MICROCATHETER, AND THREE BALT OPTIMA COILS WERE PLACED TO CONTROL THE BLEEDING. POST PROCEDURE SHOWED BLOOD IN THE BRAIN AND THE PATIENT STATUS IS UNKNOWN AT THE TIME OF REPORTING. THE REP WAS NOT INFORMED OF ANY FURTHER SURGERY PERFORMED AFTER THE CASE. ONE DAY POST PROCEDURE, THE PATIENT WAS DOING BETTER THAN EXPECTED AND ABLE TO MOVE WITHOUT ANY PARALYSIS. THE PATIENT DID STOP TALKING EARLIER ON 19JUN2025. THE SCANS WERE REPORTED AS UNCHANGED. THE PATIENT PRESENTED WITH A RUPTURED ACA WITH BLEEDING THAT HAD SUBSIDED PRIOR TO THE CASE. THE PATIENT WAS UNDER CONSCIOUS SEDATION AND WAS MOVING. THE PATIENT WAS EXPERIENCING VASOSPASM IN THE A1 SEGMENT AND THEREFORE, THE PHYSICIAN ADMINISTERED VERAPAMIL. A BALT BALLAST CATHETER 90CM, SOFIA 5F 115CM, A TRAXCESS EX WIRE, AND A WEB 17 4 X 2 WERE USED ALONG WITH THE VIA 17. THE VIA 17 WAS PARKED JUST PAST THE NECK OF THE ANEURYSM AND THE WEB WAS PREPPED IN BOWL AND PRE-TEST WAS SUCCESSFULLY PERFORMED WITH THE WDC. THE
Patient problems: Aneurysm; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Intracranial Hemorrhage
MAUDE report 22494946 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page