Skip to content

MULTILOK TABLE BASE 0302-001Recalls & reported problems

Mizuho Orthopedic Systems, Inc. · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842430111167), received between Jul 14, 2020 and Jul 14, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Failure to Eject” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Failure to Eject2

Patient problems

  • Injury2
  • Needle Stick/Puncture2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 14, 2020Adverse Event Without Identified Device or Use Problem; Failure to Eject; Adverse Event Without Identified Device or Use Problem; Failure to Eject

    STAFF WAS DISMANTLING STERILE FIELD POST PROCEDURE. WHEN DISENGAGING THE HAND TABLE TOP FROM THE BASE THE DISPOSABLE TOP PUNCTURED THROUGH THE STAFF'S GLOVES. SHE NOTICED THREE PUNCTURE HOLES THROUGH HER LEFT FIFTH FINGER. WENT TO EMERGENCY AND RECEIVED HEPATITIS C PROPHYLAXIS. SUBSEQUENT INFORMATION RECEIVED FROM THE FACILITY INDICATED THERE WAS SOME DIFFICULTY REMOVING THE (CORIN) TOP FROM THE (MIZUHO OSI) BASE WHICH CONTRIBUTED TO THE PROBLEM.

    Patient problems: Injury; Needle Stick/Puncture

    MAUDE report 10271279 ↗
  • InjuryJul 14, 2020Adverse Event Without Identified Device or Use Problem; Failure to Eject; Adverse Event Without Identified Device or Use Problem; Failure to Eject

    A STAFF MEMBER HAD DIFFICULTY RELEASING A HAND TABLE TOP FROM THE BASE AND WAS INJURED BY A SHARP EDGE ON THE TOP.

    Patient problems: Injury; Needle Stick/Puncture

    MAUDE report 10271203 ↗

FDA clearance and classification

The FDA regulates product code FWX (table, operating-room, mechanical) under 21 CFR 878.4950 as a Class I device. 21 CFR 878.4950 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the MULTILOK TABLE BASE 0302-001

Has the MULTILOK TABLE BASE 0302-001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00842430111167), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the MULTILOK TABLE BASE 0302-001?

The FDA's MAUDE database lists 2 adverse event reports for this device, received on Jul 14, 2020. In those reports, the most common specific device problem is “Failure to Eject” (2); the most common patient problems are “Injury” (2) and “Needle Stick/Puncture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MULTILOK TABLE BASE 0302-001?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MULTILOK TABLE BASE 0302-001 FDA approved?

Devices of this type (product code FWX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Mizuho Orthopedic Systems, Inc.

Who makes the MULTILOK TABLE BASE 0302-001, and is it still sold?

The FDA's device database lists Mizuho Orthopedic Systems, Inc. as the company responsible for it (brand MULTILOK TABLE BASE, model 0302-001). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Table, Operating-Room, Mechanical devices

All Table, Operating-Room, Mechanical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page