The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768011900), received between Dec 21, 2018 and Dec 21, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 21, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
FLAGGED NON-NUMERICAL PROTEIN C RESULTS WERE OBTAINED ON TWO PATIENT SAMPLES ON A SYSMEX CS-5100 SYSTEM, USING THE BERICHROM PROTEIN C REAGENT. THE FLAGGED RESULT FOR SAMPLE ID (B)(6) WAS REPORTED TO THE PHYSICIAN(S), AND IT IS UNKNOWN IF THE FLAGGED RESULT FOR SAMPLE ID (B)(6) WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE DILUTED IN A 1:4 DILUTION AND REPEATED ON THE SAME SYSTEM, USING THE SAME REAGENT LOT, RESULTING IN NUMERICAL PROTEIN C RESULTS. THE REPEAT RESULT FOR SAMPLE ID (B)(6) WAS REPORTED TO THE PHYSICIAN(S). ON (B)(6) 2018, SAMPLE ID (B)(6) WAS REPEATED AGAIN ON THE SAME SYSTEM AND REAGENT LOT, RESULTING HIGHER THAN THE REPEAT RESULT. THE CORRECT RESULT FOR EITHER PATIENT IS UNKNOWN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DIFFERENT PROTEIN C RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8189356 ↗MalfunctionDec 21, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A FLAGGED NON-NUMERICAL PROTEIN C RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A SYSMEX CS-5100 SYSTEM, USING THE BERICHROM PROTEIN C REAGENT. THE FLAGGED RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS DILUTED IN A 1:4 DILUTION AND REPEATED ON THE SAME SYSTEM, USING THE SAME REAGENT LOT, RESULTING IN A NUMERICAL PROTEIN C RESULT. THE REPEAT RESULT WAS REPORTED TO THE PHYSICIAN(S). ON (B)(6) 2018, THE SAMPLE WAS REPEATED AGAIN ON THE SAME SYSTEM AND REAGENT LOT, RESULTING HIGHER THAN THE REPEAT RESULT. THE CORRECT RESULT FOR THIS PATIENT IS UNKNOWN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DIFFERENT PROTEIN C RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8189353 ↗
Other TEST, QUALITATIVE AND QUANTITATIVE FACTOR DEFICIENCY devices
- STA® - Liatest® Free Protein S 6 00516Diagnostica Stago · TEST, QUALITATIVE AND QUANTITATIVE FACTOR DEFICIENCY1 recall0 reports
- STA® - Staclot® Protein S 00746Diagnostica Stago · TEST, QUALITATIVE AND QUANTITATIVE FACTOR DEFICIENCY1 recall0 reports
- Factor VIII Chromogenic Assay 10445729Siemens Healthcare Diagnostics Products GmbH · TEST, QUALITATIVE AND QUANTITATIVE FACTOR DEFICIENCY6 reports
- Protein C Reagent 10446185Siemens Healthcare Diagnostics Products GmbH · TEST, QUALITATIVE AND QUANTITATIVE FACTOR DEFICIENCY2 reports
- Factor VIII Antibody Screen F8SImmucor Gti Diagnostics, Inc. · Test, Qualitative And Quantitative Factor Deficiency1 report
- Protein S Ac 10445968Siemens Healthcare Diagnostics Products GmbH · TEST, QUALITATIVE AND QUANTITATIVE FACTOR DEFICIENCY1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page