Test Thrombin Reagent 10446598Recalls & reported problems
Siemens Healthcare Diagnostics Products GmbH · Determination of thrombin time in human plasma · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768012716), received between Sep 6, 2023 and Oct 24, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2023High Test Results; High Test Results
THE CUSTOMER REPORTED THAT HIGHER THROMBIN TIME (TT) RESULTS WERE OBTAINED ON FIVE PATIENT SAMPLES ON A SYSMEX CS-5100 SYSTEM USING TEST THROMBIN REAGENT LOT 02056, COMPARED TO THE RECOVERIES FROM AN ALTERNATE SYSMEX CS-5100 SYSTEM USING TEST THROMBIN REAGENT LOT 01552. THE CUSTOMER INDICATED THAT THE LOWER TT RESULTS WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DIFFERENT TT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17996820 ↗MalfunctionSep 6, 2023Low Test Results; Low Test Results
THREE FALSELY DEPRESSED THROMBIN TIME (TT) RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON TWO SYSMEX CS-5100 SYSTEMS USING TEST THROMBIN REAGENT. WHEN THE SAMPLE WAS PROCESSED INITIALLY ON THE SYSMEX CS-5100 SYSTEM (SERIAL NUMBER:(B)(6), THE SAMPLE RECOVERED HIGHER. THE INITIAL HIGHER TT RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY LOW TT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17695271 ↗
Other TEST, THROMBIN TIME devices
- BC Thrombin Reagent 10446636Siemens Healthcare Diagnostics Products GmbH · TEST, THROMBIN TIME9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.