Dimension Vista® Flex® reagent cartridge SAL 10445108Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Salicylate Flex® reagent cartridge, 80 tests · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842768015427), received between Jun 29, 2020 and Jun 29, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “High Test Results” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K061655); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 29, 2020High Test Results; High Test Results
A DISCORDANT ELEVATED SALICYLATE (SAL) RESULT WAS OBTAINED ON A PATIENT SERUM SAMPLE ON THE DIMENSION VISTA 500 SYSTEM. THE RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS PROCESSED ON THE CUSTOMER'S ALTERNATE DIMENSION VISTA SYSTEM AND AN ELEVATED RESULT WAS OBTAINED. THE SAME SAMPLE WAS PROCESSED ON AN ALTERNATE NON-SIEMENS SYSTEM AND A LOWER RESULT WAS OBTAINED. THE PATIENT WAS ADMINISTERED A MEDICATION FOR SALICYLATE POISONING. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED SALICYLATE RESULT OR THE MEDICATION TREATMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10206870 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061655. One submission often covers many models, so check the product and lot details in each record.
- Z-1660-2018 · Jun 29, 2017 · TerminatedDimension Vista Thyroxine (T4) assay; (Product Number: K6065/10445101) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
- Z-1658-2018 · Jun 29, 2017 · TerminatedDimension Vista Glucose (GLU) assay; (Product Number: K1039/10445162) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
- Z-1666-2016 · Mar 17, 2016 · TerminatedDimension Vista Assays: Uric Acid (URCA)
- Z-1665-2016 · Mar 17, 2016 · TerminatedDimension Vista Assays: Triglycerides (TRIG)
- Z-0339-2016 · Sep 30, 2015 · TerminatedDimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SMN 10445158. For the quantitative measurement of magnesium in human serum, plasma and urine on the Dimen
- Z-2175-2012 · Jul 17, 2012 · TerminatedDimension Vista(R) Vancomycin Flex(R) Reagent Cartridge (VANC, Catalog # K4086). For the quantitative measurement of vancomycin in human serum and plasma on the Dimension Vista sy
- Z-1433-2012 · Feb 17, 2012 · TerminatedDimension Vista(R) MG Flex(R) Reagent Cartridge The Dimension Vista(R) MG method is an in vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasm
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page