Dimension Vista® System FPSA Flex® Reagent Cartridge 10445091Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Free Prostate Specific Antigen Flex® reagent cartridge 80 Test · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842768016677), received between Aug 27, 2021 and Sep 3, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2021High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED FREE PROSTATE SPECIFIC ANTIGEN (FPSA) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 1500 SYSTEM. THE RESULT WAS REPORTED TO THE PHYSICIAN(S). THE RESULT WAS QUESTIONED BY THE PHYSICIAN. THE SAME SAMPLE WAS PROCESSED ON A LATER DATE AND NON-LINEAR RESULTS WERE OBTAINED. NO CORRECTED REPORT WAS ISSUED. THERE WERE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED FPSA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12423169 ↗MalfunctionAug 27, 2021High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED FREE PROSTATE SPECIFIC ANTIGEN (FPSA) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 500 SYSTEM. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME SYSTEM AND A LOWER RESULT WAS OBTAINED UPON DILUTION. THE CUSTOMER SENT OUT THE PATIENT'S SAMPLE TO A REFERENCE LABORATORY FOR TESTING USING A NON-SIEMENS METHODOLOGY. A LOWER RESULT, CONSIDERED CORRECT, WAS OBTAINED AND REPORTED. THERE WERE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED FPSA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12377242 ↗MalfunctionAug 27, 2021High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED FREE PROSTATE SPECIFIC ANTIGEN (FPSA) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 500 SYSTEM. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME DATE AND A LOWER RESULT WAS OBTAINED AND REPORTED. THE CUSTOMER SENT OUT THE PATIENT'S SAMPLE TO A REFERENCE LABORATORY FOR TESTING USING A NON-SIEMENS METHODOLOGY. A LOWER RESULT, CONSIDERED CORRECT, WAS OBTAINED AND REPORTED. THERE WERE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED FPSA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12377232 ↗
Other Test,Prostate Specific Antigen,Free,(Noncomplexed) To Distinguish Prostate Cancer From Benign Conditions devices
- Alinity 07P9321Abbott Ireland Diagnostics · Test, prostate specific antigen, free, (noncomplexed) to distinguish prostate cancer from benign conditions1 recall0 reports
- Elecsys free PSA 08828610190Roche Diagnostics GmbH · Test, prostate specific antigen, free, (noncomplexed) to distinguish prostate cancer from benign conditions25 reports
- Elecsys free PSA 07027320190Roche Diagnostics GmbH · Test, prostate specific antigen, free, (noncomplexed) to distinguish prostate cancer from benign conditions2 reports
- Elecsys free PSA 03289788160Roche Diagnostics GmbH · Test,Prostate Specific Antigen,Free,(Noncomplexed) To Distinguish Prostate Cancer From Benign Conditions2 reports
- Elecsys free PSA 08828601160Roche Diagnostics GmbH · Test, prostate specific antigen, free, (noncomplexed) to distinguish prostate cancer from benign conditions1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page