Skip to content

Dimension® CSA Flex® reagent cartridge 10285193Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Cyclosporine Flex® reagent cartridge, 80 tests · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00842768025136), received between Mar 5, 2020 and Feb 26, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results3
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2025Low Test Results; Low Test Results

    THE CUSTOMER REPORTED TO SIEMENS THAT AN ERRONEOUSLY DEPRESSED CYCLOSPORINE A (CSA) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION RXL MAX HM INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUSLY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED TWICE ON THE ORIGINAL DIMENSION RXL MAX HM INTEGRATED CHEMISTRY SYSTEM. HIGHER REPROCESSED RESULTS WERE OBTAINED AND CONSIDERED CORRECT. THE HIGHER REPROCESSED RESULT OF 116.3 NG/ML WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED CYCLOSPORINE A (CSA) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21466684 ↗
  • MalfunctionJul 1, 2024High Test Results; High Test Results

    THE CUSTOMER REPORTED ONE (1) ERRONEOUSLY ELEVATED CYCLOSPORINE A (CSA) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION® EXL¿ 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUSLY ELEVATED PATIENT RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAME SAMPLE WAS MANUALLY DILUTED AND REPROCESSED TWO TIMES ON THE SAME DIMENSION® EXL¿ 200 INTEGRATED CHEMISTRY SYSTEM. LOWER RESULTS WERE OBTAINED, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED CYCLOSPORINE A (CSA) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19642732 ↗
  • MalfunctionJun 2, 2023High Test Results; High Test Results

    TWO (2) DISCORDANT FALSELY ELEVATED CYCLOSPORINE A (CSA) RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL 200 SYSTEM. ONE OF THE FALSELY ELEVATED RESULTS WAS REPORTED TO THE PHYSICIAN(S), AND THE RESULT WAS QUESTIONED. THE SAME SAMPLE WAS REPROCESSED ON AN ALTERNATE DIMENSION EXL 200 SYSTEM, AND THE REPROCESSED RESULT WAS CONSIDERED CORRECT. A CORRECTED REPORT WAS ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CYCLOSPORINE A (CSA) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17049053 ↗
  • MalfunctionMar 5, 2020High Test Results; High Test Results

    DISCORDANT, FALSELY DEPRESSED CYCLOSPORINE A (CSA) RESULTS WERE OBTAINED ON PATIENTS' SAMPLES ON A DIMENSION EXL 200 INSTRUMENT. THE RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). AFTER A NEW FLEX WAS PUT ON-BOARD AND A RE-CALIBRATION WAS PERFORMED, EACH PATIENT SAMPLE WAS REPROCESSED ON THE SAME DATE ON THE INSTRUMENT AND A HIGHER CORRECT RESULT WAS OBTAINED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION DUE TO THE DISCORDANT, FALSELY DEPRESSED CYCLOSPORINE A RESULTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9795378 ↗

Other CYCLOSPORINE devices

All CYCLOSPORINE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page