Skip to content

CT LUCIA CT LUCIA 602.US/+21.5DRecalls & reported problems

Carl Zeiss Meditec Production, LLC · 3-piece, monofocal, hydrophobic, acrylic, intraocular lens, +21.5D · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00843045100966), received between Oct 11, 2021 and Mar 6, 2026 — classified as 18 injuries.

The most frequently reported specific device problem is “Use of Device Problem” (17 reports).

3 other recalls cover devices cleared under the same FDA submission (P100016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem17
  • Off-Label Use12
  • Adverse Event Without Identified Device or Use Problem1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Visual Impairment17
  • Eye Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2026Use of Device Problem

    IT WAS REPORTED THAT AFTER SURGERY CONCLUDED, THE DOCTOR NOTICED A BENT HAPTIC ON THE CT LUCIA 602. THE YAMANE TECHNIQUE WAS EMPLOYED DURING THE SURGERY. A SECOND CT LUCIA 602 LENS WAS IMPLANTED TWO WEEKS LATER. THE PATIENT'S CONDITION WAS GOOD AT THE ONE-WEEK POST-OPERATIVE VISIT. DURING THE PREPARATION FOR USE, NO DAMAGE WAS NOTED TO THE DEVICE.

    Patient problems: Visual Impairment

    MAUDE report 24535919 ↗
  • InjuryAug 19, 2025Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    DURING A CT LUCIA 602 LENS IMPLANTATION PROCEDURE, THE HAPTIC BECAME KINKED. THE DEVICE WAS EXPLANTED AND A BACKUP CT LUCIA 602 LENS WAS IMPLANTED WITHIN THE SAME SURGERY. THE PROCEDURE WAS PERFORMED USING A ZEISS CARTRIDGE AND INJECTOR SYSTEM. THE YAMANE TECHNIQUE WAS USED DURING THE PROCEDURE. THE AFFECTED DEVICE IS NOT RETURNABLE. NO DAMAGE WAS NOTED DURING PREPARATION OR PRIOR TO LOADING THE LENS. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Visual Impairment

    MAUDE report 22841831 ↗
  • InjuryOct 16, 2024Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT THE LENS WAS "SPINNING" IN THE EYE. THE LENS WAS EXPLANTED AND REPLACED WITH ANOTHER LENS DURING THE SAME SURGERY.

    Patient problems: Swelling/ Edema; Visual Impairment

    MAUDE report 20464724 ↗
  • InjuryMar 7, 2024Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    THE CUSTOMER REPORTED THAT THE HAPTICS WERE NOT STABLE AFTER PLACING IOL INTO THE EYE USING THE YAMANE TECHNIQUE. THE LENS HAD TO BE EXPLANTED AND REPLACED BY A DIFFERENT LENS. THE INCISION WAS ENLARGED IN ORDER TO EXPLANT THE LENS.

    Patient problems: Visual Impairment

    MAUDE report 18862354 ↗
  • InjuryMar 7, 2024Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED THAT A DEFECT IN THE OPTICAL HAPTIC WAS DISCOVERED AFTER IMPLANTATION. THE LENS HAD TO BE EXPLANTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 18855878 ↗
  • InjuryJan 9, 2024Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    CT LUCIA 602 LENS WAS REPORTED TO HAVE ROTATED AND THEN EXPLANTED. THE YAMANE TECHNIQUE AND ALCON B CARTRIDGE WAS USED FOR THE SURGERY.

    Patient problems: Visual Impairment

    MAUDE report 18477754 ↗
  • InjurySep 28, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT "A ZEISS CT LUCIA 602 WAS IMPLANTED VIA A HAPTIC FIXATED TECHNIQUE (YAMANE TECHNIQUE). AT THE END OF THE CASE, THE LENS WAS WELL CENTERED AND PLANAR WITHIN THE PUPILLARY AND VISUAL AXIS. THE NEXT MORNING ON POSTOP DAY 1, THERE WAS A DEMONSTRATING 90 DEGREE ROTATION OF THE IOL ALONG THE VERTICAL AXIS." LENS WAS EXPLANTED AND REPLACED WITH ANOTHER IOL.

    Patient problems: Visual Impairment

    MAUDE report 17836702 ↗
  • InjuryJun 6, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT A CT LUCIA 602 LENS WAS PARTIALLY IMPLANTED WHEN THE HAPTIC TWISTED. THE DOCTOR WAS ABLE TO REMOVE THE LENS WITHOUT PATIENT INJURY. NO ADVERSE EVENTS OR CLINICALLY SIGNIFICANT DELAY IN PROCEDURE REPORTED.

    Patient problems: Visual Impairment

    MAUDE report 17076070 ↗
  • InjuryMay 17, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT THE HAPTIC JUNCTION BROKE, RESULTING IN THE IOL TILTING, THE IOL WAS REMOVED AND ANOTHER LENS WAS IMPLANTED. NO ADVERSE EFFECTS REPORTED.

    Patient problems: Visual Impairment

    MAUDE report 16950749 ↗
  • InjuryMay 16, 2023Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED THAT THE LENS FLIPPED (TWISTED). THE LENS WAS REMOVED, AND ANOTHER LENS WAS IMPLANTED.

    Patient problems: Visual Impairment

    MAUDE report 16942945 ↗
  • InjuryMay 1, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT THE DOCTOR HAD A CASE WHERE THE LENS WAS POSITIONED WELL AT THE END OF THE CASE, BUT THE NEXT DAY IT HAD ROTATED A FULL 180 DEGREES. THE LENS HAS BEEN EXPLANTED.

    Patient problems: Visual Impairment

    MAUDE report 16850333 ↗
  • InjuryMar 2, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT THE IOL HAPTIC WAS DISTORTED OR TWISTED WITHIN THE HAPTIC JUNCTION AFTER INITIAL SUCCESSFUL SCLERAL FIXATION, CAUSING SIGNIFICANT LENS TILT. THE LENS WAS EXPLANTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 16470500 ↗
  • InjuryFeb 6, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT AFTER IMPLANTING A CT LUCIA 602 THE IOL TILTED.

    Patient problems: Visual Impairment

    MAUDE report 16309422 ↗
  • InjuryJan 23, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT AFTER IMPLANTING A CT LUCIA 602 LENS THE IOL TILTED POD1. LENS WAS EXPLANTED ANOTHER LENS WAS IMPLANTED. PATIENT WAS NOT INJURED, NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 16230719 ↗
  • InjuryJan 9, 2023Off-Label Use; Use of Device Problem; Off-Label Use; Use of Device Problem

    IT WAS REPORTED THAT AFTER TRYING TO MANIPULATE THE TRAILING HAPTIC FOR THE YAMANE TECHNIQUE THE HAPTIC DISINSERTED FROM THE OPTIC WHEN GRASPED WITH THE INTRAOCULAR FORCEPS. THE LENS WAS REMOVED, AND ANOTHER LENS WAS IMPLANTED. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN PROCEDURE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 16128048 ↗
  • InjuryFeb 2, 2022Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT AFTER A CT LUCIA 602 WAS IMPLANTED IT WAS NOTICED THERE WAS A KINK IN THE HAPTIC. THUS, THE LENS WAS REMOVED AND A BACKUP LENS WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN PROCEDURE REPORTED.

    Patient problems: Visual Impairment

    MAUDE report 13423813 ↗
  • InjuryDec 3, 2021Off-Label Use; Use of Device Problem; Improper or Incorrect Procedure or Method; Off-Label Use; Use of Device Problem; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT A CT LUCIA 602 WAS IMPLANTED USING THE SCLERAL FIXATION TECHNIQUE. THE PATIENT COULDN'T TOLERATE THE PAIN CAUSED BY THE SCLERAL FIXATION TECHNIQUE; THUS, THE DOCTOR DECIDED TO REMOVE THE LENS AND IMPLANT AN ANTERIOR CHAMBER LENS. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED AND NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Eye Pain

    MAUDE report 12926799 ↗
  • InjuryOct 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER A CT LUCIA 602 LENS WAS INSERTED INTO A PATIENT'S EYE IT WAS NOTED THAT THE HAPTIC WAS BROKEN. THE IOL WAS REMOVED FROM THE EYE SUCCESSFULLY WITHOUT PATIENT INJURY. SPARE LENS WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN PROCEDURE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Visual Impairment

    MAUDE report 12606192 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1309-2015 · Feb 13, 2015 · Terminated
    Intraocular Lens (IOL), Part No. EC-3 PAL, Diopter: 17.5, 18.0, 18.5, 19.0, 19.5, 21.5, 23.5, 9.0, 22.5, 23.0, 21.0, 24.0. The device is indicated for primary implantation in the
  • Z-1077-2012 · Aug 2, 2011 · Terminated
    Precision Aspheric Lens (PAL) Intraocular Lens Model: EC-3 PAL, Serial Number: 1169381106, 1169391106, 1169421106 The device is indicated for primary implantation in the capsular
  • Z-1076-2012 · Aug 2, 2011 · Terminated
    Intraocular Lens (IOL). Model: EC-3, Serial Number: 1169371106, 1169401106, 1169471106 The device is indicated for primary implantation in the capsular bag of the eye for the visu

Other Intraocular Lens devices

All Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page