Skip to content

RelayPro Thoracic Stent-Graft System 28-M4-40-145-40URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576149816), received between Dec 19, 2024 and Dec 19, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance1
  • Insufficient Device Problem Information1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1
  • Rupture1
  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 19, 2024Difficult to Advance; Insufficient Device Problem Information; Difficult to Advance; Insufficient Device Problem Information

    "PATIENTS WAS SCHEDULED FOR TEVAR PROCEDURE ON (B)(6)2024. PROCEDURE WAS GOING WELL UNTIL "S" THORACIC AORTIC WOULD NOT STRAIGHTEN OUT WITH 2 BILATERAL LUNDQUIST WIRES. THE FIRST DEVICE WAS ABLE TO ADVANCE THE TRANSFERS AORTA BUT NO FURTHER DUE TO THE ANGULATION OF THE THORACIC AORTA. THE DEVICE WAS DEPLOYED 4CM DISTAL TO LSA WITHOUT INCIDENT. THE 2ND DEVICE NEEDED TO BE ADVANCED PROXIMALLY TO EXISTING STENT GRAFT JUST DISTAL TO LSA. THE DECISION WAS TO USE BODY-FLOSS FROM THE RIGHT BRACHIAL TO RIGHT FEMORAL TO CREATE ENOUGH TENSION ON THE WIRE TO RAIL THE DEVICE. THIS WAS DONE SUCCESSFULLY BUT THE DISTAL ANGIOGRAM SHOWED AN AORTIC DISSECTION. THIS DECISION WAS TO LEAVE IT ALONE AS THE DISSECTION DID NOT EXTEND ABOVE EXISTING STENT GRAFT OUR BELOW EXISTING ABDOMINAL STENT GRAFT. ON (B)(6)2024 AT 1330 THE PHYSICIAN CALLED ME TO INFORM ME THAT THE PATIENT'S HEMOGLOBIN HAD DECREASED, AND CT WAS ORDERED AND SHOWS A THORACIC RUPTURE AT THE LEVEL OF THE VISCERAL VESSELS. PHYSICIAN HAS DECIDED TO NOT TREAT AS THE PATIENT IS 85 YEARS OLD AND HAS OTHER CO-MOBILITIES. (PHYSICIAN SENT ME VIDEO OF CT THAT SHOWS RUPTURE)." PATIENT OUTCOME: "PATIENT CURRENTLY IS STABLE, BUT HOSPICE HAS BEEN CONT

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions; Rupture; Vascular Dissection

    MAUDE report 20982027 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page