Skip to content

RelayPro Thoracic Stent-Graft System 28-M4-30-200-30URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576149892), received between Oct 23, 2025 and Oct 23, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Mechanical Problem; Difficult to Advance; Mechanical Problem; Difficult to Advance

    "A 30X200MM RELAY PRO BARE STENT WAS INTRODUCED THROUGH THE PATIENTS RIGHT COMMON "FEMORAL" "ARTERY". THE DEVICE WAS POSITIONED IN THE "APPROPRIATE" LOCATION OF THE "DESCENDING" AORTA PROXIMAL TO THE CELIAC ARTERY. THE PHYSICIAN "VERIFIED" THE DEVICE WAS IN POSITION 1 ON THE CONTROLLER AND BEGAN TO ROLL THE BLACK DEPLOYMENT GRIP CLOCKWISE (AWAY FROM HIM) TO ADVANCE THE INNER SHEATH INTO THE APPROPRIATE POSITION. THE INNER SHEATH DID NOT RESPOND AND REMAINED IN THE PRIMARY SHEATH. THE PHYSICIAN THEN PRESSED THE BLACK DISENGAGEMENT BUTTON AND ADVANCED THE INNER SHEATH MANUALLY. WHEN THE PROXIMAL END OF THE INNER SHEATH WAS IN POSITION, DISTAL TO THE LSA, WE NOTICED THAT THERE WAS STILL ONE STENT REMAINING IN THE PROXIMAL SHEATH. HOWEVER, THE DISENGAGEMENT BUTTON WAS MOVED ALL THE WAY FORWARD MEETING THE CONTROLLER AND COULD NOT BE "ADVANCED" ANY FARTHER. THE PHYSICIAN TRIED ROLLING THE DEPLOYMENT GRIP AGAIN AND STILL NO RESPONSE FROM THE INNER SHEATH. WE CALLED SEVERAL ASSOCIATES IN THE FIELD AS WELL AS ENGINEERING FOR SOME BAIL OUTS AND WORK AROUNDS. IN THE END, THE PHYSICIAN WAS ABLE TO "ACCORDION, TRAIN-WRECK", THE PRIMARY SHEATH ON ITSELF BY HAVING THE FELLOW PUSH THE DEVICE FORW

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368116 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates