Skip to content

RelayPro Thoracic Stent-Graft System 28-M4-36-190-36URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00843576149922), received between Jan 15, 2025 and Feb 27, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Endoleak” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Endoleak3
  • Failure to Seal3
  • Malposition of Device1

Patient problems

  • Clinical Endoleak3
  • Aneurysm expansion1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 27, 2026Device Endoleak

    "SITE DOCUMENTED IN ELECTRONIC DATA CAPTURE (EDC) SYSTEM THAT THERE WAS NO DEVICE DEFICIENCY DURING OR AFTER THE INDEX OR EXTENSION PROCEDURE AND NO ENDOLEAKS PRESENT AT THE END OF EITHER PROCEDURE. ON POST OPERATIVE DAY 722 (INDEX)/284(EXTENSION), PATIENT (B)(6) HAD A CT SCAN ON (B)(6) 2026 WHICH SHOWED MULTIPLE STABLE UNCLASSIFIED ENDOLEAKS IN THE PROXIMAL STENT AND MIDPORTION OF THE STENT WITH OVERALL STABLE AGGREGATE DIAMETER MEASURING UP TO 7.2CM. (B)(6) 2025, CTA REVEALED INTERVAL GROWTH OF THE PROXIMAL DESCENDING AORTA MEASURING A MAXIMUM OF 7.1 CM IN DIAMETER. FOLLOW-UP IMAGING IN (B)(6) 2025 SHOWED ONGOING LEAK FROM THE DISTAL GRAFT AND ENLA.... (CHARACTER LIMIT REACHED IN THE EDC SYSTEM). TREATMENT OF THE EVENT ON (B)(6) 2026 INCLUDED EXTENSION TEVAR VIA RIGHT CFA CUTDOWN. THE VASCULAR PROCEDURE INCLUDED PERCUTANEOUS ACCESS OF THE RIGHT FEMORAL ARTERY WITH A 7-FRENCH SHEATH, A RIGHT ILIAC ANGIOGRAM, AND A STENTING OF THE RIGHT COMMON ILIAC ARTERY AND EXTERNAL ILIAC A.... (CHARACTER LIMIT REACHED IN EDC SYSTEM). TYPE 1B ENDOLEAK OF A THORACIC ENDOVASCULAR AORTIC REPAIR GRAFT DONE PREVIOUSLY AFTER THE PATIENT HAS AN EXTENSIVE HISTORY OF TYPE A AORTIC DISSECTION AND EXTENSIV

    Patient problems: Clinical Endoleak

    MAUDE report 24473294 ↗
  • InjuryApr 17, 2025Malposition of Device; Failure to Seal; Device Endoleak; Malposition of Device; Failure to Seal; Device Endoleak

    "ENDOLEAK TYPE 1A: DURING DEPLOYMENT OF THE PROXIMAL STENT-GRAFT, AS DEPLOYMENT PROGRESSED TO THE SECOND STENT, THE STENT-GRAFT WAS SUBJECTED TO A DISTAL FORCE AND MIGRATED DISTALLY. THE STENT-GRAFT LANDED MORE DISTALLY THAN PLANNED, RESULTING IN INADEQUATE SEALING AT THE PROXIMAL NECK, CAUSING A TYPE 1A ENDOLEAK. NO ADDITIONAL STENT-GRAFT WAS PLACED TO TREAT THE ENDOLEAK DURING THIS OPERATION AS HEALTH INSURANCE WOULD NOT COVER THE PLACEMENT OF A THIRD STENT-GRAFT ON THE SAME DAY. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER TREATMENT. AN ADDITIONAL STENT-GRAFT WILL BE IMPLANTED AT A LATER DATE. OPERATION TYPE: TEVAR. BLOOD LOSS: UNKNOWN. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY. PRE-CASE PLAN AVAILABLE: SCHEMA. ADDITIONAL INFORMATION WILL BE OBTAINED UPON REQUEST. DELIVERY SYSTEM WAS DISCARDED BY USER. (TC#(B)(4))". PATIENT OUTCOME: "THERE WAS NO HARM TO THE PATIENT'S HEALTH."

    Patient problems: Clinical Endoleak

    MAUDE report 21860321 ↗
  • InjuryFeb 24, 2025Failure to Seal; Failure to Seal

    "CT IMAGING SHOWED STENT GRAFT INDUCED NEW ENTRY TEAR WITH RAPIDLY GROWING DESCENDING DISTAL AORTIC ANEURYSM. PI DETERMINED THE EVENT RELATED TO THE DEVICE (MAIN BODY). CELIAC ARTERY STENOSIS AND FEVAR UP TO ZONE 6 DISTALLY WAS THE ACTION TAKE REGARDING THIS EVENT. EVENT HAS BEEN RESOLVED WITHOUT SEQUALAE. NOTE THAT AN EVENT INVOLVING THESE DEVICES WAS PREVIOUSLY REPORTED AND INVESTIGATED UNDER COMPLAINT NUMBER (B)(4)." PATIENT OUTCOME: "PI MONITORING EVENT; WILL REVIEW IMAGES AT NEXT F/U."

    Patient problems: Aneurysm expansion

    MAUDE report 21451893 ↗
  • InjuryJan 15, 2025Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak

    "TYPE 1A ENDOLEAK AS THE PROXIMAL NECK OF THE ANEURYSM WAS SHORT, THE RELAYPRO STENT-GRAFT WAS PLACED PARTIALLY COVERING THE LSA. TYPE 1A ENDOLEAK WAS OBSERVED ON THE LESSER CURVATURE SIDE. THE PATIENT WAS PLACED UNDER OBSERVATION BECAUSE THE ENDOLEAK WAS SMALL. OPERATION TYPE: TEVAR BLOOD LOSS: UNKNOWN. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY PRE-CASE PLAN AVAILABLE ADDITIONAL INFORMATION WILL BE OBTAINED UPON REQUEST (TC#BM 241202508)" PATIENT OUTCOME: "NO HARM TO THE PATIENT'S HEALTH".

    Patient problems: Clinical Endoleak

    MAUDE report 21162193 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page