RelayPro Thoracic Stent-Graft System 28-M4-46-200-46URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of RelayPro Thoracic Stent-Graft System (127)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576149977), received between Aug 4, 2025 and Aug 4, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Difficult or Delayed Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 20, 2022Class IIOpen, Classified
The product may be shorter than packaging indicates
Z-0997-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult or Delayed Separation1
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 4, 2025Malposition of Device; Difficult or Delayed Separation; Malposition of Device; Difficult or Delayed Separation
"UNABLE TO DEPLOY THE STENT GRAFT: THE RELAY PRO STENT-GRAFT WAS USED FOR EMERGENCY 2-DEBRANCHING TEVAR IN A RUPTURE CASE. AFTER A C-TAG STENT-GRAFT (37 - 10, GORE) WAS IMPLANTED ON THE DISTAL SIDE, THE RELAY PRO STENT-GRAFT WAS IMPLANTED STACKED ON THE FIRST STENT-GRAFT FROM ZONE 1. THE STENT-GRAFT WAS DEPLOYED AS PER THE PROCEDURE AND THE APEX HOLDER KNOB WAS SLID TOWARD THE DISTAL SIDE, BUT THE PROXIMAL BARE STENT WAS NOT RELEASED FROM THE APEX HOLDER. THE APEX HOLDER CAPTURING THE PROXIMAL BARE STENT WAS NOT SLID DISTALLY. WHILE THE APEX HOLDER KNOB WAS BEING PUSHED AND PULLED TO APPLY FORCE TO THE ENTIRE DEVICE PROXIMALLY AND THE HANDLE BODY WAS BEING ROTATED HALF A TURN CLOCKWISE AND ANTICLOCKWISE, THE PROXIMAL BARE STENT WAS RELEASED FROM THE APEX HOLDER WITHOUT REALIZING IT. THE DELIVERY SYSTEM WAS THEN REMOVED FROM THE PATIENT. PROBABLY BECAUSE OF SUCH BEHAVIOR, THE STENT-GRAFT WAS IMPLANTED LOWER THAN THE PLANNED POSITION. NO LEAKS WERE FOUND AS FAR AS THE PHYSICIAN CHECKED. OPERATION TYPE: EMERGENCY 2-DEBRANCHING TEVAR BLOOD LOSS: AMOUNT UNKNOWN . ANCILLARY DEVICE USED: C-TAG STENT-GRAFT (37 - 10, GORE) NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY NO PRE-CASE PLAN AVAILABLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22705909 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.
- Z-2160-2026 · Apr 22, 2026 · Open, ClassifiedRelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
- Z-1383-2024 · Feb 22, 2024 · CompletedRELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page