Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576150027), received between Sep 18, 2024 and Dec 26, 2025 — classified as 1 death, 1 injury.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 20, 2022Class IIOpen, Classified
The product may be shorter than packaging indicates
Z-0997-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Advance1
- Failure to Advance1
- Retraction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Perforation of Vessels1
- Rupture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathDec 26, 2025Retraction Problem; Adverse Event Without Identified Device or Use Problem; Retraction Problem; Adverse Event Without Identified Device or Use Problem
"UNABLE TO DEPLOY THE STENT GRAFT / RUPTURE / DEATH: AFTER AN ABDOMINAL AORTIC REPLACEMENT WITH BYPASS TO FOUR ABDOMINAL BRANCHES, TEVAR WAS PERFORMED AS A SINGLE-STAGE PROCEDURE. AFTER A RELAY PRO STENT-GRAFT WAS IMPLANTED IN THE THORACIC DESCENDING AORTA, THIS RELAY PRO STENT-GRAFT WAS ATTEMPTED TO BE DELIVERED CENTRALLY. HOWEVER, IT TOOK TIME TO DELIVER THE DEVICE TO THE TARGET SITE. DURING THIS TIME, BLOOD PRESSURE DECREASED, INDICATING A RUPTURE. ONCE THE DEVICE REACHED THE TARGET SITE, DEPLOYMENT WAS ATTEMPTED. DEPLOYMENT OF THE STENT-GRAFT WAS STARTED WITH THE CONTROLLER IN THE "2" POSITION. HOWEVER, DESPITE PULLING THE DEPLOYMENT GRIP FULLY TO THE PROXIMAL SIDE, THE INNER SHEATH DID NOT PULL DOWN, AND THE STENT-GRAFT DID NOT DEPLOY. DUE TO THE DEPLOYMENT FAILURE, THE DEVICE WAS REMOVED FROM THE PATIENT. THE DEVICE WAS REPLACED WITH A TAG STENT-GRAFT (GORE), WHICH WAS SUCCESSFULLY IMPLANTED. HOWEVER, THE DESCENDING AORTA RUPTURED, AND THE PATIENT EVENTUALLY DIED DURING THE PROCEDURE. AS THE PATIENT'S CONDITION WAS POOR, THE DOCTOR COMMENTED THE THERE WAS NO CAUSAL RELATIONSHIP BETWEEN THE DEVICE AND DEATH. DOCTOR'S COMMENT: THE RELAY PRO STENT-GRAFT WAS USED DURING THE PROCE
Patient problems: No Clinical Signs, Symptoms or Conditions; Rupture
MAUDE report 23898483 ↗InjurySep 18, 2024Failure to Advance; Difficult to Advance; Failure to Advance; Difficult to Advance
"UNABLE TO ADVANCE THE INNER SHEATH/BLEEDING: THE RELAYPRO WAS INSERTED FROM THE RIGHT FEMORAL ARTERY FOR TEVAR AND ADVANCED TO THE IMPLANTATION SITE. THE INNER SHEATH WAS ADVANCED CENTRALLY WITH THE CONTROLLER IN POSITION 1, BUT ADVANCEMENT OF THE INNER SHEATH WAS SLOW AND DID NOT FULLY COME OUT OF THE OUTER SHEATH. THEREFORE, THE RELAYPRO WAS WITHDRAWN AND REMOVED FROM THE PATIENT. ANOTHER RELAYPRO DEVICE OF DIFFERENT SPECIFICATIONS WAS USED TO CONTINUE THE PROCEDURE, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. WHEN THE RELAYPRO WAS BEING WITHDRAWN, THE INTIMA NEAR THE RIGHT FEMORAL ARTERY WAS SLIGHTLY PEELED OFF. ANGIOGRAPHY WAS PERFORMED TO CHECK FOR DAMAGE AND THE PHYSICIAN DETERMINED THAT IT WAS MINOR, BUT A LIFESTREAM COVERED STENT (BECTON, DICKINSON AND COMPANY) WAS IMPLANTED IN THE DAMAGED AREA FOR SAFETY. PHYSICIAN'S COMMENT: BLOOD LOSS WAS APPROXIMATELY 100~200ML DURING THE PROCEDURE. IT WAS GOOD THAT THE RELAYPRO WAS SAFELY REMOVED, AND THE PATIENT WAS NOT AFFECTED TOO MUCH. WE WOULD LIKE YOU TO CLARIFY THE CAUSE OF WHY SUCH AN EVENT OCCURRED. OPERATION TYPE: TEVAR BLOOD LOSS: APPROXIMATELY 100~200ML ANCILLARY DEVICE: EGOIST STIFF GUIDEWIRE (MEDICO'S HIRATA) IMAGE AVA
Patient problems: Perforation of Vessels
MAUDE report 20249121 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.
- Z-2160-2026 · Apr 22, 2026 · Open, ClassifiedRelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
- Z-1383-2024 · Feb 22, 2024 · CompletedRELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page