RelayPro Thoracic Stent-Graft System 28-M4-34-145-30URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576150140), received between Oct 3, 2024 and Oct 30, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0997-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Endoleak1
  • Difficult or Delayed Separation1
  • Failure to Eject1
  • Separation Failure1

Patient problems

  • Clinical Endoleak1
  • Perforation of Vessels1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 30, 2024Separation Failure; Failure to Eject; Difficult or Delayed Separation; Separation Failure; Failure to Eject; Difficult or Delayed Separation

    "APEX HOLDER COULD NOT BE PULLED/DELIVERY SYSTEM TIP, APEX HOLDER AND OUTER CONTROL TUBE WERE LEFT IN THE PATIENT/DISSECTION: BACKGROUND OF THE PATIENT: TWO VALIANT (APPROX. SIZE 34, MEDTRONIC) WERE IMPLANTED CENTRALLY AND THE RELAYPRO WAS TO BE IMPLANTED PERIPHERALLY TO THE VALIANT DEVICES. THERE WAS A SEVERE CURVATURE IN THE TARGET VESSEL. IMPLANTATION PLAN: -THE RELAY PRO (28-M4-34-145-30U) WAS TO BE IMPLANTED USING THE PULL-THROUGH TECHNIQUE. -WITHOUT PLACING THE OUTER SHEATH AT THE VESSEL CURVATURE TO AVOID DAMAGE FROM THE TIP OF THE OUTER SHEATH, THE INNER SHEATH WAS TO BE ADVANCED GRADUALLY FROM THE PROXIMAL TO THE CURVATURE TO ADVANCE THE DELIVERY SYSTEM UNDER THE CONTROL OF A GUIDEWIRE FOR THE PULL-THROUGH TECHNIQUE. IMPLANTATION: AS PER THE ABOVE PLAN, THE DELIVERY SYSTEM COULD BE ADVANCED TO THE TARGET LANDING ZONE BY CONTROLLING THE GUIDEWIRE FOR THE PULL-THROUGH TECHNIQUE. RELEASE OF THE STENT GRAFT: -THE STENT GRAFT WAS SUCCESSFULLY DEPLOYED WITH THE CONTROLLER IN POSITION 2. -TO RELEASE THE BARE STENT OF THE STENT GRAFT BY PULLING THE APEX HOLDER (INDICATING 3), THE APEX HOLDER WAS ATTEMPTED TO BE PULLED BUT COULD NOT BE PULLED. THE APEX HOLDER WAS PULLED FORCEFULLY

    Patient problems: Perforation of Vessels

    MAUDE report 20565974 ↗
  • InjuryOct 3, 2024Device Endoleak; Device Endoleak

    "ENDOLEAK (TYPE UNKNOWN): THE RELAYPRO STENT-GRAFT WAS IMPLANTED FROM ZONE 2. THE FINAL ANGIOGRAPHY CONFIRMED NO LEAKAGE, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. AFTER THAT, HOWEVER, CONTRAST-ENHANCED CT REVEALED AN ENDOLEAK (TYPE UNKNOWN). ON (B)(6) 2024, ADDITIONAL TEVAR WAS PERFORMED. A RELAYPRO (28-M4-34-145-30U), WHICH HAS COMPLETELY DIFFERENT SPECIFICATIONS, WAS IMPLANTED OVER THE STENT-GRAFT CONCERNED THAT HAD ALREADY BEEN IMPLANTED. THE PHYSICIAN DETERMINED THAT THE ENDOLEAK HAD DISAPPEARED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. OPERATION TYPE: TEVAR BLOOD LOSS: UNKNOWN. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY. NO PRE-CASE PLAN AVAILABLE DUE TO HOSPITAL POLICY. NO ADDITIONAL INFORMATION AVAILABLE DUE TO HOSPITAL POLICY. (B)(4)" PATIENT OUTCOME - "HEALTH DAMAGE TO THE PATIENT IS UNKNOWN."

    Patient problems: Clinical Endoleak

    MAUDE report 20363476 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices