Skip to content

RelayPro Thoracic Stent-Graft System 28-M4-38-145-34URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150164), received between May 15, 2026 and May 15, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0997-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Advance1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2026Failure to Advance; Unintended Movement

    "BROKEN: THE RELAY PRO STENT GRAFT WAS USED FOR ONE-DEBRANCHING TEVAR. AFTER CONSTRUCTING THE DEBRANCH, THE DELIVERY SYSTEM WAS INSERTED THROUGH THE RIGHT COMMON FEMORAL ARTERY AND ADVANCED TO THE LEVEL OF THE AORTIC VALVE. AFTER ENSURING THAT THE CONTROLLER WAS IN THE "1" POSITION, THE DEPLOYMENT GRIP WAS ROTATED CLOCKWISE. THE PHYSICIAN FELT SOMETHING UNUSUAL BECAUSE HE HEARD A CRACKING SOUND AT THE BEGINNING OF THE ROTATION, BUT HE CONTINUED TO ROTATE THE GRIP. THE INNER SHEATH WAS ADVANCED UNTIL THE DEPLOYMENT GRIP ASSEMBLY PASSED THE WHITE ARROW MARKER WITH THE CONTROLLER IN THE "1" POSITION. HOWEVER, UNDER FLUOROSCOPY, THE INNER SHEATH DID NOT APPEAR TO ADVANCE OUT OF THE OUTER SHEATH TIP. UPON LOOKING AT THE COMMON FEMORAL ARTERY, THE PHYSICIAN FOUND THAT THE OUTER SHEATH HAD DETACHED FROM THE HANDLE BODY. THEREFORE, THE DELIVERY SYSTEM WAS REMOVED FROM THE PATIENT. ANOTHER RELAY PRO STENT GRAFT WITH THE SAME SPECIFICATIONS (AR-B3814534, LOT#2503110143) WAS INSERTED. THE STENT GRAFT WAS DEPLOYED AND IMPLANTED WITHOUT ISSUE. SUBSEQUENTLY, THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PHYSICIAN COMMENTED THAT THE OUTER SHEATH COULD NOT HAVE PROPERLY FIT WITH THE HANDLE BODY. O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25192082 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page