RelayPro Thoracic Stent-Graft System 28-M4-40-145-36URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of RelayPro Thoracic Stent-Graft System (127)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150171), received between Aug 13, 2025 and Aug 13, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Failure to Advance” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance1
- Inadequacy of Device Shape and/or Size1
Patient problems
- Perforation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 13, 2025Inadequacy of Device Shape and/or Size; Failure to Advance; Inadequacy of Device Shape and/or Size; Failure to Advance
"BROKEN: THE RELAY PRO STENT GRAFT WAS USED FOR A PATIENT WITH POOR ACCESS AND A SEVERELY BENT DESCENDING AORTA. AFTER 2-DEBRANCHING PROCEDURE, TEVAR WAS PERFORMED ON THE AORTIC ARCH. DUE TO THE DIFFERENCE IN VESSEL DIAMETER BETWEEN THE CENTER AND PERIPHERAL SIDES, A RELAY PRO STENT GRAFT (28-N4-34-109-34U, LOT # 2503180106) WAS FIRST INSERTED ON THE PERIPHERAL SIDE USING THE PULL-THROUGH TECHNIQUE. AT THAT TIME, THE STENT GRAFT WAS DELIVERED OVER THE GUIDEWIRE WHILE THE GUIDEWIRE WAS MOVED FORWARD AND BACKWARD TO RELEASE SLACK, BUT IT WAS DIFFICULT TO DELIVER THE STENT GRAFT. THE STENT GRAFT WAS THEN SOMEHOW DELIVERED TO THE INTENDED SITE AND DEPLOYED. THE STENT GRAFT WAS SUCCESSFULLY IMPLANTED. THEN, THE OUTER SHEATH OF THE RELAY PRO STENT GRAFT CONCERNED WAS ADVANCED TO ZONE 1 USING THE PULL-THROUGH TECHNIQUE. HOWEVER, THE STENT GRAFT COULD NOT PASS THROUGH THE BEND OF THE IMPLANTED RELAY PRO STENT GRAFT, WHICH PREVENTED THE STENT GRAFT FROM BEING DELIVERED. AT A CERTAIN POINT, THE PHYSICIAN ROTATED THE DEPLOYMENT GRIP WITH THE CONTROLLER IN THE "1" POSITION IN AN ATTEMPT TO ADVANCE THE INNER SHEATH. HOWEVER, ONLY THE GRIP ROTATED; THE INNER SHEATH DID NOT ADVANCE. ASSUMING THAT
Patient problems: Perforation
MAUDE report 22796528 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.
- Z-2160-2026 · Apr 22, 2026 · Open, ClassifiedRelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
- Z-1383-2024 · Feb 22, 2024 · CompletedRELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page