RelayPro Thoracic Stent-Graft System 28-M4-36-190-32URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of RelayPro Thoracic Stent-Graft System (127)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150256), received between Nov 25, 2025 and Nov 25, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size1
- Patient-Device Incompatibility1
Patient problems
- Vascular Dissection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 25, 2025Inadequacy of Device Shape and/or Size; Patient-Device Incompatibility; Inadequacy of Device Shape and/or Size; Patient-Device Incompatibility
"DISSECTION: ON AUGUST 12, THE PATIENT WITH A DISSECTING AORTIC ANEURYSM UNDERWENT TEVAR AND HAD THE RELAY PRO STENT-GRAFT (BS, 36 MM) IMPLANTED IN THE PROXIMAL SEALING PORTION (VESSEL DIAMETERS: 30 MM - 27 MM - 33 MM). ON THURSDAY EVENING, OCTOBER 30, RETROGRADE TYPE A AORTIC DISSECTION (RTAD) OCCURRED AT THE BARE-STENT ON THE LESSER CURVATURE SIDE OF THE RELAY PRO STENT-GRAFT, REQUIRING RE-INTERVENTION. ON FRIDAY, OCTOBER 31, TOTAL ARCH REPLACEMENT (TAR) WAS PERFORMED. A THORAFLEX HYBRID WITH A STENT DIAMETER OF 30 MM WAS IMPLANTED, AND THE ENTRY WAS RESECTED. THE SURGERY WAS SUCCESSFULLY COMPLETED, AND INFORMATION ON POSTOPERATIVE PROGRESS WILL BE OBTAINED LATER. PHYSICIAN'S COMMENTS ON CAUSALITY: THIS EVENT MAY HAVE BEEN DUE TO THE LARGE DIAMETER OF THE STENT-GRAFT AND THE SELECTION OF THE RELAY PRO BARE-STENT TYPE. PHYSICIAN'S COMMENT: THE LARGE DIAMETER OF THE STENT-GRAFT AND THE PLAN TO USE THE RELAY PRO BARE-STENT TYPE FOR THE DISSECTING AORTIC ANEURYSM MAY NOT HAVE BEEN UNDESIRABLE. OPERATION TYPE: TEVAR BLOOD LOSS: UNKNOWN. IMAGE WILL BE ABLE TO BE OBTAINED. PRE-CASE PLAN WILL BE ABLE TO BE OBTAINED. ADDITIONAL INFORMATION WILL BE ABLE TO BE OBTAINED. (TC#(B)(4))". PATIEN
Patient problems: Vascular Dissection
MAUDE report 23645269 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.
- Z-2160-2026 · Apr 22, 2026 · Open, ClassifiedRelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
- Z-1383-2024 · Feb 22, 2024 · CompletedRELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page