RelayPro Thoracic Stent-Graft System 28-M4-40-195-36URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of RelayPro Thoracic Stent-Graft System (127)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576150270), received between Oct 16, 2024 and Jan 15, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Endoleak” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Endoleak2
- Failure to Seal1
Patient problems
- Clinical Endoleak2
- Neuropathy1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 15, 2025Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak
"DURING THE EARLY FOLLOW-UP A TYPE 1B ENDOLEAK WAS FOUND ON THE IMAGING. CONTRAST MATERIAL IN EXCLUDED SAC AT THE DISTAL ASPECT OF THE STENT, WHICH THE PI DEEMED RELATED TO MAIN BODY OF DEVICE. SUBJECT ALSO HAD LOWER EXTREMITY PARESTHESIA. WEAKNESS, TINGLING, AND NUMBNESS IN BOTH LEGS FROM LOW BLOOD PRESSURE." PATIENT OUTCOME: "PI MONITORING EVENT; AND DETERMINE IF INTERVENTION IS NEEDED."
Patient problems: Clinical Endoleak; Neuropathy
MAUDE report 21162261 ↗InjuryOct 16, 2024Device Endoleak; Device Endoleak
"ENDOLEAK (TYPE IIIA): STENT-GRAFT IMPLEMENTATION WAS PERFORMED TO TREAT A DESCENDING AORTIC ANEURYSM. A STENT-GRAFT (28-N4-34-209-34U) WAS FIRST IMPLEMENTED ON THE DISTAL SIDE AND THE STENT-GRAFT (28-M4-40-195-36U) WAS THEN IMPLEMENTED ON THE PROXIMAL SIDE. POST DILATATION WAS PERFORMED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ONE WEEK AFTER THE PROCEDURE, HOWEVER, CONTRAST-ENHANCED CT REVEALED A TYPE IIIA ENDOLEAK. SINCE THE PATIENT IS OLD, THE PATIENT IS BEING FOLLOWED UP. IF THE ANEURYSM TENDS TO ENLARGE, ADDITIONAL TREATMENT WILL BE SCHEDULED. OPERATION TYPE: STENT GRAFT IMPLANTATION. BLOOD LOSS: UNKNOWN. ANCILLARY DEVICE USED: 28-N4-34-209-34U. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY. PRE-CASE PLAN AVAILABLE (SEE THE ATTACHED SCHEMA.) ADDITIONAL INFORMATION WILL BE ABLE TO BE OBTAINED. (TC# (B)(6))." PATIENT OUTCOME: "NO HEALTH DAMAGE TO THE PATIENT."
Patient problems: Clinical Endoleak
MAUDE report 20465866 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.
- Z-2160-2026 · Apr 22, 2026 · Open, ClassifiedRelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
- Z-1383-2024 · Feb 22, 2024 · CompletedRELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page