Skip to content

RelayPro Thoracic Stent-Graft System 28-M4-44-200-40URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150294), received between Jan 7, 2025 and Jan 7, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Endoleak1
  • Failure to Seal1
  • Insufficient Device Problem Information1

Patient problems

  • Clinical Endoleak1
  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 7, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Failure to Seal; Device Endoleak; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Failure to Seal; Device Endoleak

    "ENDOLEAK (TYPE1A): ONLY TWO-DEBRANCHING PROCEDURE WAS PERFORMED THE DAY BEFORE TEVAR AND ONLY TEVAR WAS PERFORMED ON (B)(6). AS PER THE STANDARD PROCEDURE, 28-N4-36-109-36U (2310160270) WAS IMPLANTED PERIPHERALLY AND THE RELAYPRO CONCERNED WAS ATTEMPTED TO BE IMPLANTED CENTRALLY FROM JUST BELOW THE BCA. THE RELAYPRO WAS IMPLANTED, BUT IT COULD NOT BE SUFFICIENTLY CONFIRMED WHETHER THE BCA WAS COMPLETELY COVERED WITH THE RELAYPRO OR NOT. THE FINAL ANGIOGRAPHY SHOWED NO LEAKS, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. HOWEVER, ON (B)(6), A LEAK WAS CONFIRMED ON POST-OPERATIVE ANGIOGRAPHY. THE PHYSICIAN ALSO SUSPECTED A TYPE 3B ENDOLEAK, BUT WHEN THE PHYSICIAN CHECKED THE ACTUAL IMPLANTATION SITE, THE RELAYPRO WAS IMPLANTED NOT JUST BELOW THE BCA, BUT WHERE IT COVERED HALF OF THE LCCA. THERE IS A POSSIBILITY THAT AN ADDITIONAL TEVAR WILL BE PERFORMED, BUT THIS IS STILL UNDECIDED AT THIS TIME. OPERATION TYPE: TEVAR AFTER 2 DEBRANCHING. ANCILLARY DEVICE: 28-N4-36-109-36U (2310160270). BLOOD LOSS: SMALL AMOUNT. IMAGE AVAILABLE UPON REQUEST. PRE-CASE PLAN AVAILABLE: SCHEMA ATTACHED. ADDITIONAL INFORMATION AVAILABLE UPON REQUEST. (TC#:(B)(4)". PATIENT OUTCOME: "NO HEALTH DAMAGE TO THE

    Patient problems: Clinical Endoleak; Insufficient Information

    MAUDE report 21091868 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates