Skip to content

RelayPro Thoracic Stent-Graft System 28-M4-40-250-36URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150379), received between Jan 29, 2025 and Jan 29, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0997-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Advance1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2025Failure to Advance; Difficult to Advance; Failure to Advance; Difficult to Advance

    "TYPE OF PROCEDURE PERFORMED: TEVAR. WHILE DEPLOYING THE DOCTOR NEVER CAME OUT OF POSITION #1. LANDED SHORT AND PROCEEDED TO TURN THE DEPLOYMENT KNOB TO ADVANCE THE INNER SHEATH. WAS RESPONDING BUT NOT GREAT. IT WAS STILL SOMEWHAT IN THE OUTER SHEATH BUT STILL NEEDED TO BE ADVANCED TO GET TO TARGET AT THE LSA. THE DOCTOR PUSHED ENTIRE SYSTEM AS A UNIT TO THE TARGET AREA WITH MINIMAL EFFORT. WE NOTICED THAT THE OUTER SHEATH WAS MORE THAN 3/4 COVERING THE IMPLANT WHICH WAS STILL IN THE INNER SHEATH AND NOT DEPLOYED YET. WE THOUGHT TO REVERSE DEPLOY BUT THE WAS REALLY NO ROOM FOR THIS AS THE MECHANICAL ADV WAS HUBBED AGAINST THE CONTROLLER AND THE OUTER SHEATH WAS HUBBED ON THE PATIENT. WHEN LOOKING IN THE MAIN BODY CHANNEL IT APPEARED THAT EVERYTHING WAS JAMMED UP. WE DECIDED TO REMOVE THE ENTIRE UNIT WHICH WAS UN-EVENTFUL AND OPEN THE BACKUP (WHICH WAS NBS, NOT THE BARE) AND THE PROCEDURE WAS PERFECT AFTER THAT. PATIENT IDENTIFIER: (B)(6). PATIENT OUTCOME: "THERE WAS NO HARM, THE PATIENT IS STABLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21264943 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates