Skip to content

RelayPro Thoracic Stent-Graft System 28-N4-26-104-26URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150430), received between Nov 6, 2024 and Nov 6, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult or Delayed Separation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0998-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult or Delayed Separation1
  • Ejection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2024Ejection Problem; Difficult or Delayed Separation; Ejection Problem; Difficult or Delayed Separation

    "DEFECTIVE: THE RELAYPRO CONCERNED AND A RELAYPRO (28-N4-34-154-30U, LOT#2405280110) WERE IMPLANTED STACKED IN A THORACIC AORTIC DISSECTION CASE. DURING IMPLANTATION OF THE RELAYPRO CONCERNED, THE STENT GRAFT WAS DEPLOYED WITH THE CONTROLLER IN POSITION 2 WITHOUT ANY PROBLEMS. HOWEVER, WHEN THE APEX HOLDER KNOB (MARKED 3) WAS PULLED BACK PROXIMALLY, THE CLASP WAS NOT RELEASED. ALTHOUGH THE APEX HOLDER KNOB COULD BE PULLED, THE APEX HOLDER KNOB AND THE APEX HOLDER WERE NOT INTERLOCKED, AND THE APEX HOLDER COULD NOT BE PULLED. BY USING PEAN FORCEPS TO GRASP AND PULL THE OUTER CONTROL TUBE (GREEN TUBE) IN THE SLOT AT THE PROXIMAL END OF THE STAINLESS-STEEL ROD, AND THE APEX HOLDER COULD BE PULLED, AND THE STENT GRAFT WAS SUCCESSFULLY RELEASED. SUBSEQUENTLY, THE PROCEDURE WAS SUCCESSFULLY COMPLETED. OPERATION TYPE: TEVAR. ANCILLARY DEVICE: 28-N4-34-154-30U (LOT#2405280110). BLOOD LOSS: UNKNOWN. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY. NO PRE-CASE PLAN AVAILABLE DUE TO HOSPITAL POLICY. NO ADDITIONAL INFORMATION AVAILABLE DUE TO HOSPITAL POLICY (TC# (B)(4))." PATIENT OUTCOME: "NO HEALTH DAMAGE TO THE PATIENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20624872 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2160-2026 · Apr 22, 2026 · Open, Classified
    RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page